Skip to the content

FDA Medical Device Recalls (openFDA)

Eizo Corporation: RadiForce RX360, model 0FTD2346 - Product Usage: is a color LCD monitor indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. The display is not intended for mammography.

Recall number
Z-2294-2020
Recalling firm
Eizo Corporation
Product
RadiForce RX360, model 0FTD2346 - Product Usage: is a color LCD monitor indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. The display is not intended for mammography.
Status
Open, Classified
Initiated
2019-07-30
Posted
not on file
Terminated
not on file
Reason
Under certain conditions, a marble pattern infrequently appears on the monitor.
Root cause
Device Design
Action
EIZO Japan announced the recall on their website on 7/30/2019. EIZO U.S. began contacting their consignees via email and phone (one as early as 8/7/2019, some on 10/18/2019, but the majority on 11/26/2019) and began updating consignee monitors with the firmware. The notification requested the consignee contact EIZO U.S. for assistance with the firmware update.
Quantity
1,466 monitors dist. to the U.S.
Distribution
Worldwide distribution - US Nationwide distribution including in the states of Arizona, Arkansas, California, Colorado, Connecticut, Delaware, Florida, Georgia, Hawaii, Illinois, Indiana, Kansas, Louisiana, Maine, Massachusetts, Michigan, Missouri, Montana, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, Ohio, Oklahoma, Pennsylvania, South Dakota, Tennessee, Texas, Virginia, Washington, and Wisconsin. Distribution was also made to Puerto Rico. There was no government/military consignees. The countries of Canada, Australia, Austria, Belgium, Brazil, China, Czech Republic, Egypt, France, Germany Greece, Hong Kong, India, Israel, Italy, Japan, Korea, Lebanon, Netherlands, Pakistan, Poland, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, and United Kingdom.
Lot, serial or model codes
RadiForce RX360, 1.0001 or earlier, UDI 04995047053279, 04995047053293.
510(k) numbers
["K182591"]
PMA numbers
not on file
Product code
PGY
Firm FEI
3000237660
Firm city
Hakusan
Firm state
not on file
Firm country
Japan

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.