Skip to the content

FDA Medical Device Recalls (openFDA)

Kico Knee Innovation Company: Arvis Hip & Knee 2.0 Instrument Set, REF: IN-21200 containing one of the following components: IN-10500, Impactor V-Block Assembly; IN-10600 , Tracker B; IN-16400, Tracker A; IN-16100, Tracker C

Recall number
Z-2293-2026
Recalling firm
Kico Knee Innovation Company
Product
Arvis Hip & Knee 2.0 Instrument Set, REF: IN-21200 containing one of the following components: IN-10500, Impactor V-Block Assembly; IN-10600 , Tracker B; IN-16400, Tracker A; IN-16100, Tracker C
Status
Open, Classified
Initiated
2026-04-23
Posted
2026-06-02
Terminated
not on file
Reason
The Impactor V-Block Assembly and Trackers A, B, C and E may exhibit signs of early wear and/or corrosion on the exposed magnet face.
Root cause
Device Design
Action
On April 27, 2026 Kico Knee Innovation issued a Urgent: Medical Device Recall to affected consignees via E-Mail. Kico ask consignees to take the following actions: 1. Instruments can continue to be used as is, given none of the failure modes are exhibited. 2. This notice needs to be passed on to all those who need to be aware within your organization or where the potentially affected devices have been transferred. 3. Please retain a copy of this letter, place a copy where devices are stored, and keep a copy of the acknowledgement form for your files. 4. Complete the attached Acknowledgement Form immediately and return it email to Alexander.Borchardt@enovis.com 5. Instruments should always be visually inspected prior to use as stated in the Instructions for Use.
Quantity
124 kits
Distribution
Worldwide - US Nationwide distribution in the states of MS, CA, IA, IN, MI, CO, MT, FL, IL, NY, AL, WI, NE, AZ, MN, VT, NH, MD, RI, MA, NJ, PA, KS, WA, ID, OR, TN, VA, PR and the countries of Australia, France, Germany, Italy.
Lot, serial or model codes
All Lots/serial Numbers; UDI: IN-28000: 00810180350020; IN-10500: (01)00810180350686; IN-10600 (01)00810180350693; IN-16400:(01)00810180350525; IN-16100: (01)00810180350563;
510(k) numbers
["K203115"]
PMA numbers
not on file
Product code
OLO
Firm FEI
3013552417
Firm city
Pymble
Firm state
not on file
Firm country
Australia

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register