Siemens Medical Solutions, USA, Inc: syngo Dynamics version 9.0, Siemens Medical Solutions, USA, Inc., Ann Arbor, MI. The syngo Dynamics is a picture archiving and communications system intended for acceptance, transfer, display, storage, archiving and manipulation of digital medical images, including quantification and report generation. It is not intended to be used for the reading of mammography images.
- Recall number
- Z-2293-2011
- Recalling firm
- Siemens Medical Solutions, USA, Inc
- Product
- syngo Dynamics version 9.0, Siemens Medical Solutions, USA, Inc., Ann Arbor, MI. The syngo Dynamics is a picture archiving and communications system intended for acceptance, transfer, display, storage, archiving and manipulation of digital medical images, including quantification and report generation. It is not intended to be used for the reading of mammography images.
- Status
- Terminated
- Initiated
- 2011-03-24
- Posted
- 2011-05-25
- Terminated
- 2012-07-10
- Reason
- If one or more graphs are in a table on the report, and if any of those graphs are deleted, it may lead to reports failing to save and/or marking the study as read might leave the study in an unread state or might mark the study as read but leave the report with a "Preliminary" watermark displayed.
- Root cause
- Software design
- Action
- The firm, SIEMENS, sent a "Customer Safety Advisory Notice" letter dated March 24, 2011 to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to place graphs in sections and not tables in the report; avoid deleting graphs from the reports in run-time and include this safety advisory notice in their operator's manual where it should remain until the service patch is applied. SIEMENS will be issuing a service patch to fix the defect. The letter states the patch will be available in the next few weeks. If you have any questions, call +1 (734) 205-2400.
- Quantity
- 29 devices
- Distribution
- Nationwide distribution: USA including states of: CA, CT, FL, ID, MD, MI, MN, NJ, NY and PA.
- Lot, serial or model codes
- 10091602, 10091604
- 510(k) numbers
- ["K102150"]
- PMA numbers
- not on file
- Product code
- LLZ
- Firm FEI
- 3003304172
- Firm city
- Ann Arbor
- Firm state
- MI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28