Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions, USA, Inc: syngo Dynamics version 9.0, Siemens Medical Solutions, USA, Inc., Ann Arbor, MI. The syngo Dynamics is a picture archiving and communications system intended for acceptance, transfer, display, storage, archiving and manipulation of digital medical images, including quantification and report generation. It is not intended to be used for the reading of mammography images.

Recall number
Z-2293-2011
Recalling firm
Siemens Medical Solutions, USA, Inc
Product
syngo Dynamics version 9.0, Siemens Medical Solutions, USA, Inc., Ann Arbor, MI. The syngo Dynamics is a picture archiving and communications system intended for acceptance, transfer, display, storage, archiving and manipulation of digital medical images, including quantification and report generation. It is not intended to be used for the reading of mammography images.
Status
Terminated
Initiated
2011-03-24
Posted
2011-05-25
Terminated
2012-07-10
Reason
If one or more graphs are in a table on the report, and if any of those graphs are deleted, it may lead to reports failing to save and/or marking the study as read might leave the study in an unread state or might mark the study as read but leave the report with a "Preliminary" watermark displayed.
Root cause
Software design
Action
The firm, SIEMENS, sent a "Customer Safety Advisory Notice" letter dated March 24, 2011 to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to place graphs in sections and not tables in the report; avoid deleting graphs from the reports in run-time and include this safety advisory notice in their operator's manual where it should remain until the service patch is applied. SIEMENS will be issuing a service patch to fix the defect. The letter states the patch will be available in the next few weeks. If you have any questions, call +1 (734) 205-2400.
Quantity
29 devices
Distribution
Nationwide distribution: USA including states of: CA, CT, FL, ID, MD, MI, MN, NJ, NY and PA.
Lot, serial or model codes
10091602, 10091604
510(k) numbers
["K102150"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
3003304172
Firm city
Ann Arbor
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register