Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer Gmbh: Avenir Mller Stem 4 Standard Avenir Mller Stem 6 Lateral Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Intended Use: " Advanced wear of the joint due to degenerative, post-traumatic or rheumatic diseases. " Failed previous hip surgery including joint reconstruction (osteotomy), arthrodesis, hemiarthroplasty or total hip replacement (THR). " Acute traumatic fracture of the femoral head or neck " Avascular necrosis of the femoral head. " Avenir Moller Stems are for cementless use only.

Recall number
Z-2292-2016
Recalling firm
Zimmer Gmbh
Product
Avenir Mller Stem 4 Standard Avenir Mller Stem 6 Lateral Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Intended Use: " Advanced wear of the joint due to degenerative, post-traumatic or rheumatic diseases. " Failed previous hip surgery including joint reconstruction (osteotomy), arthrodesis, hemiarthroplasty or total hip replacement (THR). " Acute traumatic fracture of the femoral head or neck " Avascular necrosis of the femoral head. " Avenir Moller Stems are for cementless use only.
Status
Terminated
Initiated
2016-06-30
Posted
2016-07-26
Terminated
2017-03-27
Reason
Product mix-up. The Avenir Mller Stem 6 lateral uncemented might be placed in the packaging of the Avenir Mller Stem 4 standard uncemented and vice versa. Potential for delay in surgery, and/or potential for use of incorrect implant which may in turn lead dismetry in patient and possibly the need for revision.
Root cause
Labeling mix-ups
Action
Zimmer Biomet sent an Urgent Medical Device Recall letter dated June 30, 2016 to all affected customers via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers were instructed to review the notification immediately and ensure affected personnel are aware of the contents and assist Zimmer Biomet sales representative with the quarantine of any affected device. Your Zimmer Biomet sales representative will remove any affected device. Customers were also instructed to complete the Certification of Acknowledgement form and return to corporatequality.postmarket@zimmerbiomet.com. Customers with further questions or concerns were instructed to contact the customer call center at 1-800-348-2759 between 8:00 am and 8:00 pm EST. For questions regarding this recall call 800-348-2759.
Quantity
2
Distribution
Nationwide Distribution to OH and NM
Lot, serial or model codes
Item Name: Avenir M¿ller Stem 4 Standard Item Number: 01.06010.004 Lot Number: 4022860  Item Name: Avenir M¿ller Stem 6 Lateral Item Number: 01.06010.106 Lot Number: 4023094
510(k) numbers
["K123392"]
PMA numbers
not on file
Product code
LZO
Firm FEI
3005233524
Firm city
Winterthur
Firm state
not on file
Firm country
Switzerland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register