Skip to the content

FDA Medical Device Recalls (openFDA)

Wavelight AG: WaveLight FS200 Femtosecond Laser System Alcon Product Code 8065990714 WaveLight Model 1025 Software Version: Version Blue, Service Pack 2. The WaveLight FS 200 Laser System is an ophthalmic surgical laser indicated for use in refractive surgery for LASIK flap cutting, lamellar and perforating keratoplasty, and intrastromal surgery for implants.

Recall number
Z-2292-2011
Recalling firm
Wavelight AG
Product
WaveLight FS200 Femtosecond Laser System Alcon Product Code 8065990714 WaveLight Model 1025 Software Version: Version Blue, Service Pack 2. The WaveLight FS 200 Laser System is an ophthalmic surgical laser indicated for use in refractive surgery for LASIK flap cutting, lamellar and perforating keratoplasty, and intrastromal surgery for implants.
Status
Terminated
Initiated
2011-04-18
Posted
2011-05-25
Terminated
2011-09-23
Reason
The software calculating the patterns were found to has a rounding error for certain parameter ranges. These rounding errors could result in an incomplete cut in the donor or patient tissue.
Root cause
Other
Action
WaveLight AG sent an URGENT MEDICAL DEVICE CORRECTION letter April 18, 2011 to the affected customers. The letter identified the product, problem and the action to be taken by the customer. The letter advised the customers to discontinue keratoplasty cuts with the WaveLight FS200 until the new version can be installed. Alcon will contact you to schedule a visit of a service engineer at your site, to disable the keratoplasty function in your device, until a new software version can be installed to fix this problem. If you have experienced a product problem or adverse event related to this medical device correction, please contact Alcon Medical Safety at 1-800-561-6466, option 3. Customers were requested to complete the attached reply form to acknowledge receipt of this notification and fax it to ( 817) 916-9087.
Quantity
55 units
Distribution
Worldwide Distribution - USA including states of TX, KS, NC, CA, FL, OH, SC, MD, VA and NY and countries of Switzerland, Germany, Greece, United Kingdom, Ireland, Spain, Italy, Australia, India, Malaysia, France, Finland, Saudi Arabia, Czech Republic, Poland, China and South Korea and Iran.
Lot, serial or model codes
Alcon Product Code 8065990714 WaveLight Model 1025 Software Version: Version Blue, Service Pack 2
510(k) numbers
["K101006"]
PMA numbers
not on file
Product code
HNO
Firm FEI
3003288808
Firm city
Pressath
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register