FDA Medical Device Recalls (openFDA)
Tornier S.A.S.: The Aequalis Pyrocarbon Humeral Heads. Shoulder implant component. Sold OUS only: a) Catalog # DWH039 - Diameter 39 mm Height 14 mm Eccentric LOW (1.5) Material PyC/CoCr, b) Catalog # DWH041 - Diameter 41 m m Height 15 mm Eccentric LOW (1.5) Material PyC/CoCr, c) Catalog # DWH043 - Diameter 43 mm Height 16 mm Eccentric LOW (1.5) Material PyC/CoCr, d) Catalog # DWH046 - Diameter 46 mm Height 17 mm Eccentric LOW (1.5) Material PyC/CoCr, e) Catalog # DWH048 - Diameter 48 mm Height 18 mm Eccentric LOW (1.5) Material PyC/CoCr, f) Catalog # DWH050 - Diameter 50 mm Height 16 mm Eccentric LOW (1.5) Material PyC/CoCr, g) Catalog # DWH051 - Diameter 50 mm Height 19 mm Eccentric LOW (1.5) Material PyC/CoCr, h) Catalog # DWH052 - Diameter 52 mm Height 19 mm Eccentric LOW (1.5) Material PyC/CoCr, i) Catalog # - DWH053 - Diameter 52 mm Height 23 mm Eccentric LOW (1.5) Material PyC/CoCr, j) Catalog # DWH054 - Diameter 54 mm Height 23 mm Eccentric LOW (1.5) Material PyC/CoCr, k) Catalog # DWH139 - Diameter 39 mm Height 14 mm Eccentric HIGH (3.5) Material PyC/CoCr, l) Catalog # DWH141 - Diameter 41 mm Height 15 mm Eccentric HIGH (3.5) Material PyC/CoCr, m) Catalog # DWH143 - Diameter 43 mm Height 16 mm Eccentric HIGH (3.5) Material PyC/CoCr, n) Catalog # DWH146 - 46 mm Height 17 mm Eccentric HIGH (4) Material PyC/CoCr, o) Catalog # DWH148, - Diameter 48 mm Height 18 mm Eccentric HIGH (4) Material PyC/CoCr, p) Catalog # DWH150 - Diameter 50 mm Height 16 mm Eccentric HIGH (4) Material PyC/CoCr, q) Catalog # DWH151 - Diameter 50 mm Height 19 mm Eccentric HIGH (4) Material PyC/CoCr, r) Catalog # DWH152 - Diameter 52 mm Height 19 mm Eccentric HIGH (4) Material PyC/CoCr, s) Catalog # DWH153 - Diameter 52 mm Height 23 mm Eccentric HIGH (4) Material PyC/CoCr, t) Catalog # DWH154 - Diameter 54 mm Height 23 mm Eccentric HIGH (4) Material PyC/CoCr
- Recall number
- Z-2291-2021
- Recalling firm
- Tornier S.A.S.
- Product
- The Aequalis Pyrocarbon Humeral Heads. Shoulder implant component. Sold OUS only: a) Catalog # DWH039 - Diameter 39 mm Height 14 mm Eccentric LOW (1.5) Material PyC/CoCr, b) Catalog # DWH041 - Diameter 41 m m Height 15 mm Eccentric LOW (1.5) Material PyC/CoCr, c) Catalog # DWH043 - Diameter 43 mm Height 16 mm Eccentric LOW (1.5) Material PyC/CoCr, d) Catalog # DWH046 - Diameter 46 mm Height 17 mm Eccentric LOW (1.5) Material PyC/CoCr, e) Catalog # DWH048 - Diameter 48 mm Height 18 mm Eccentric LOW (1.5) Material PyC/CoCr, f) Catalog # DWH050 - Diameter 50 mm Height 16 mm Eccentric LOW (1.5) Material PyC/CoCr, g) Catalog # DWH051 - Diameter 50 mm Height 19 mm Eccentric LOW (1.5) Material PyC/CoCr, h) Catalog # DWH052 - Diameter 52 mm Height 19 mm Eccentric LOW (1.5) Material PyC/CoCr, i) Catalog # - DWH053 - Diameter 52 mm Height 23 mm Eccentric LOW (1.5) Material PyC/CoCr, j) Catalog # DWH054 - Diameter 54 mm Height 23 mm Eccentric LOW (1.5) Material PyC/CoCr, k) Catalog # DWH139 - Diameter 39 mm Height 14 mm Eccentric HIGH (3.5) Material PyC/CoCr, l) Catalog # DWH141 - Diameter 41 mm Height 15 mm Eccentric HIGH (3.5) Material PyC/CoCr, m) Catalog # DWH143 - Diameter 43 mm Height 16 mm Eccentric HIGH (3.5) Material PyC/CoCr, n) Catalog # DWH146 - 46 mm Height 17 mm Eccentric HIGH (4) Material PyC/CoCr, o) Catalog # DWH148, - Diameter 48 mm Height 18 mm Eccentric HIGH (4) Material PyC/CoCr, p) Catalog # DWH150 - Diameter 50 mm Height 16 mm Eccentric HIGH (4) Material PyC/CoCr, q) Catalog # DWH151 - Diameter 50 mm Height 19 mm Eccentric HIGH (4) Material PyC/CoCr, r) Catalog # DWH152 - Diameter 52 mm Height 19 mm Eccentric HIGH (4) Material PyC/CoCr, s) Catalog # DWH153 - Diameter 52 mm Height 23 mm Eccentric HIGH (4) Material PyC/CoCr, t) Catalog # DWH154 - Diameter 54 mm Height 23 mm Eccentric HIGH (4) Material PyC/CoCr,
- Status
- Open, Classified
- Initiated
- 2021-06-25
- Posted
- not on file
- Terminated
- not on file
- Reason
- Potential patient exposure to the graphite substrate.
- Root cause
- Device Design
- Action
- On June 25, 2021 Tornier sent the following notification by letter: With regards to cases where this product has already been implanted, we suggest that physicians check for the potential presence of metallic bodies in the shoulder joint and adapt patient monitoring, as appropriate, when metallic bodies are present. Otherwise, we suggest that physicians continue the monitoring of patients per their standard clinical protocols. Actions needed 1. Please circulate this Field Safety Notice to all surgeons using AEQUALIS" PYROCARBON Humeral Head Implant 2. Maintain awareness of this communication internally until all required actions have been completed within your facility. 3. Return the enclosed business reply form by email to FieldAction@wright.com within 7 days to confirm receipt of this notification. 4. Inform Tornier SAS if any of the subject devices have been distributed to other organizations. If so, provide contact details so Tornier SAS can inform the recipients appropriately. 5. Please inform Tornier SAS of any adverse event and/or report them to the Competent Authorities in accordance with current regulations and in compliance with MEDDEV 2.12 If you have any questions or concerns, please contact Customer Service +1 888 494 7950.
- Quantity
- 6631 units
- Distribution
- CA, CO, CT, FL, KS, LA, MD, MI, MN, MO, NC, OR, PA, TX and WI
- Lot, serial or model codes
- All Lot Codes
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- QHE
- Firm FEI
- 3000931034
- Firm city
- Montbonnot St Martin
- Firm state
- not on file
- Firm country
- France
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28