Skip to the content

FDA Medical Device Recalls (openFDA)

Biocare Medical, LLC: 4plus Streptavidin HRP Label, Conjugated Streptavidin Horseradish Peroxidase, Detection Component, Catalog Number: HP604H

Recall number
Z-2291-2020
Recalling firm
Biocare Medical, LLC
Product
4plus Streptavidin HRP Label, Conjugated Streptavidin Horseradish Peroxidase, Detection Component, Catalog Number: HP604H
Status
Terminated
Initiated
2019-10-10
Posted
not on file
Terminated
2020-12-02
Reason
Firm has identified product has no staining. If used, may result in invalid test results when used with controls in the clinical setting. Positive control would observe no staining. The clinical impact may result in the consumption of patient tissue sample and a delay in diagnosis.
Root cause
Pending
Action
On 10/10/2019, the firm notified affected customers via posting and email of "Medical Device Recall Notification" letter. indicating: (1) Please immediately discontinue use and distribution of the identified affected lot number. Ship affected products using the UPS or FedEx account provided to the firm. A product replacement will be sent to the customer at no charge. (2) If you may have further distributed this product, please identify those customers and notify them at once of this product recall. Your notification to your customers should include a copy of this recall notification letter. (3) Please complete and return the enclosed Return Response Form by October 25, 2019 via fax via fax at 1- 925-603-8080 (Attn: Regulatory Affairs). The completed form may also be emailed to the regulatory team. Recalling firm telephone contact: 1-800-799-9499.
Quantity
8
Distribution
US: TX, MA, TN, MO, IN
Lot, serial or model codes
Lot Number: 041619-4, Expiration Date: 2020/10
510(k) numbers
not on file
PMA numbers
not on file
Product code
NJT
Firm FEI
3004140393
Firm city
Pacheco
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register