Skip to the content

FDA Medical Device Recalls (openFDA)

Johnson & Johnson Surgical Vision, Inc.: TECNIS Eyhance IOL, REF: ICB00, STERILE.EO, Rx Only

Recall number
Z-2289-2023
Recalling firm
Johnson & Johnson Surgical Vision, Inc.
Product
TECNIS Eyhance IOL, REF: ICB00, STERILE.EO, Rx Only
Status
Open, Classified
Initiated
2023-07-06
Posted
2023-08-03
Terminated
not on file
Reason
Their is a potential that intraocular lenses (IOLs) were manufactured with unintended toric fiducial marks.
Root cause
Under Investigation by firm
Action
On June 07, 20223, Johnson & Johnson Vison issued an "Urgent: Medical Device Recall" notice to affected consignees. Consignees were asked to take the following actions: 1. Identify if any of your inventory contains TECNIS IOLs from the list on pages 5-6 (see page 4 for label examples). 2. Remove all affected TECNIS IOLs from your inventory. No other TECNIS IOLs are affected by this recall. 3. Complete the attached Customer Reply Form (on page 3). We require this information for reconciliation purposes with regulatory agencies, even if you have no inventory. 4. Follow the specific instructions included in the letter "If you have product to be returned" or "If you do not have product to be returned" 5. Share this notice with anyone within your organization that needs to be informed and to any organization where the potentially affected products have been transferred. 6. If you have product complaints or adverse events to report regarding the use of the affected TECNIS IOLs, please inform Johnson & Johnson Vision. If you do report a complaint, please provide the TECNIS IOL serial number and, if a patient was involved, the date of surgery, a description of the event and patient outcome. 7. In addition, you may report any adverse events experienced or quality issues with the use of the TECNIS IOLs to the FDA's MedWatch Adverse Event Reporting program by phone at 1-800-332-1088, by fax at 1-800-332-0178, by mail at MedWatch, FDA, 5600 Fishers Lane, Rockville, MD 20852-9787, or on the MedWatch Web site at www.fda.gov/medwatch/report.htm.
Quantity
61 units
Distribution
International distribution to the countries of China, India, Korea.
Lot, serial or model codes
UDI-DI: N/A/ Serial Numbers: 8046642236, 8046652236, 8046662236, 8046672236, 8046682236, 8046692236, 8046702236, 8046712236, 8046722236, 8046732236, 8046742236, 8046752236, 8046762236, 8046772236, 8046782236, 8046792236, 8046802236, 8046812236, 8046822236, 8046832236, 8046842236, 8046852236, 8046862236, 8046872236, 8046882236, 8046892236, 8046902236, 8046912236, 8046922236, 8046932236, 8046942236, 8046952236, 8046962236, 8046972236, 8046982236, 8046992236, 8047002236, 8047012236, 8047022236, 8047032236, 8047042236, 8047052236, 8047062236, 8047072236, 8047082236, 8047102236, 8047112236, 8047122236, 8047132236, 8047142236, 8047152236, 8047162236, 8047172236, 8047182236, 8047192236, 8047202236, 8047212236, 8047222236, 8047232236, 8047242236, 8047252236
510(k) numbers
not on file
PMA numbers
["P980040S117"]
Product code
HQL
Firm FEI
3012236936
Firm city
Irvine
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register