Zimmer Inc.: Zimmer ITST" INTERTROCHANTERIC/SUBTROCHANTERIC FIXATION LOCKING BOLT EXTRACTOR - NONSTERILE Zimmer, Warsaw, IN 46580, U.S.A. The ITST Intertrochanteric/Subtrochanteric Fixation Locking Bolt Extractor instrument is a reusable surgical instrument used to separate the femoral nail from the locking bolt after nail implantation/fixation in trauma surgeries
- Recall number
- Z-2289-2010
- Recalling firm
- Zimmer Inc.
- Product
- Zimmer ITST" INTERTROCHANTERIC/SUBTROCHANTERIC FIXATION LOCKING BOLT EXTRACTOR - NONSTERILE Zimmer, Warsaw, IN 46580, U.S.A. The ITST Intertrochanteric/Subtrochanteric Fixation Locking Bolt Extractor instrument is a reusable surgical instrument used to separate the femoral nail from the locking bolt after nail implantation/fixation in trauma surgeries
- Status
- Terminated
- Initiated
- 2010-07-05
- Posted
- 2010-08-26
- Terminated
- 2011-02-25
- Reason
- Instruments were manufactured with the incorrect grade of steel, increasing the potential for fracture during use.
- Root cause
- Nonconforming Material/Component
- Action
- Zimmer sent "Urgent: Device Removal" letters dated July 8, 2010 to consignees identifying the potential for fracture of the device. Consignees were instructed to inventory affected product and record on the Inventory Return Certification Form, to document users of the affected device on the User Facility/HCP Form, and return both forms by fax or email. Affected product is to be returned to Zimmer with a copy of the Inventory Return Certification Form. Returned product will be destroyed. Each of the affected consignees will be provided with a new instrument, manufactured with the correct 17-4 stainless steel. Consignees can contact Zimmer at 800-613-6131 or 574-372-4463
- Quantity
- 20 individual units
- Distribution
- Worldwide Distribution in the countries of US, Canada, Spain, Germany, and Great Britain.
- Lot, serial or model codes
- Catalog No. 00-2258-051-01; Lot No. 53951300; Manufactured 8/8/2002
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- LXH
- Firm FEI
- 1000220733
- Firm city
- Warsaw
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28