Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer Inc.: Zimmer ITST" INTERTROCHANTERIC/SUBTROCHANTERIC FIXATION LOCKING BOLT EXTRACTOR - NONSTERILE Zimmer, Warsaw, IN 46580, U.S.A. The ITST Intertrochanteric/Subtrochanteric Fixation Locking Bolt Extractor instrument is a reusable surgical instrument used to separate the femoral nail from the locking bolt after nail implantation/fixation in trauma surgeries

Recall number
Z-2289-2010
Recalling firm
Zimmer Inc.
Product
Zimmer ITST" INTERTROCHANTERIC/SUBTROCHANTERIC FIXATION LOCKING BOLT EXTRACTOR - NONSTERILE Zimmer, Warsaw, IN 46580, U.S.A. The ITST Intertrochanteric/Subtrochanteric Fixation Locking Bolt Extractor instrument is a reusable surgical instrument used to separate the femoral nail from the locking bolt after nail implantation/fixation in trauma surgeries
Status
Terminated
Initiated
2010-07-05
Posted
2010-08-26
Terminated
2011-02-25
Reason
Instruments were manufactured with the incorrect grade of steel, increasing the potential for fracture during use.
Root cause
Nonconforming Material/Component
Action
Zimmer sent "Urgent: Device Removal" letters dated July 8, 2010 to consignees identifying the potential for fracture of the device. Consignees were instructed to inventory affected product and record on the Inventory Return Certification Form, to document users of the affected device on the User Facility/HCP Form, and return both forms by fax or email. Affected product is to be returned to Zimmer with a copy of the Inventory Return Certification Form. Returned product will be destroyed. Each of the affected consignees will be provided with a new instrument, manufactured with the correct 17-4 stainless steel. Consignees can contact Zimmer at 800-613-6131 or 574-372-4463
Quantity
20 individual units
Distribution
Worldwide Distribution in the countries of US, Canada, Spain, Germany, and Great Britain.
Lot, serial or model codes
Catalog No. 00-2258-051-01; Lot No. 53951300; Manufactured 8/8/2002
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register