Spark Biomedical Inc: Sparrow Ascent Patient Controller, REF's: 11-110-K, 11-310-K, 11-320-K, 11-321-K, 11-310-K-28, 11-320-K-28, 11-321-K-28
- Recall number
- Z-2287-2025
- Recalling firm
- Spark Biomedical Inc
- Product
- Sparrow Ascent Patient Controller, REF's: 11-110-K, 11-310-K, 11-320-K, 11-321-K, 11-310-K-28, 11-320-K-28, 11-321-K-28
- Status
- Open, Classified
- Initiated
- 2025-07-02
- Posted
- 2025-08-11
- Terminated
- not on file
- Reason
- Due to manufacturing issue, neurostimulator device may experience "cable disconnect" error message and interruption to stimulation output.
- Root cause
- Process control
- Action
- On July 2, 2025, Spark Biomedical initiated a Urgent Medical Device Recall via phone. During the phone conversation, Spark Biomedical will ask consignees to take the following actions: 1. Spark Biomedical will replace all affected00 units at no cost. Even if the units appear to be working. 2. Current devices may continue to be used until they are replaced. 3. Please perform a check of the cable to patient controller connection before use. If this connection is loose or moves, please discontinue use of that patient controller device and quarantine that device until the replacement arrives. 4. Please notify all appropriate personnel at your facility that need to know about this recall.
- Quantity
- 78 units
- Distribution
- US Nationwide distribution.
- Lot, serial or model codes
- Lot numbers: V2408xxxx and runs through V2519xxxx UDI: 11-110/00850052017033; 11-310-K/00860005396948; 11-320-K/00860005396986; 11-321-K/00850052017163; 11-310-K-28/00850052017088; 11-320-K-28/00850052017071; 11-321-K-28/00850052017170
- 510(k) numbers
- ["K230796"]
- PMA numbers
- not on file
- Product code
- PZR
- Firm FEI
- 3017987352
- Firm city
- Richmond
- Firm state
- TX
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28