FDA Medical Device Recalls (openFDA)

Alcon Research LLC: Acrysof, IQ IOL w/UltraSert System, AcrySof IQ ASPHERIC IOL, 20.5 D, SP ACRYLIC FOLDABLE LENS, w/UltraSert DELIVERY SYSTEM, UV w/BLUE LIGHT FILTER. 13.0mm. LENGTH, 6.0mm ANTERIOR ASYMMETRIC BICONVEX OPTIC, PLANAR HAPTICS.

Recall number
Z-2287-2020
Recalling firm
Alcon Research LLC
Product
Acrysof, IQ IOL w/UltraSert System, AcrySof IQ ASPHERIC IOL, 20.5 D, SP ACRYLIC FOLDABLE LENS, w/UltraSert DELIVERY SYSTEM, UV w/BLUE LIGHT FILTER. 13.0mm. LENGTH, 6.0mm ANTERIOR ASYMMETRIC BICONVEX OPTIC, PLANAR HAPTICS.
Status
Terminated
Initiated
2020-04-17
Posted
not on file
Terminated
2021-02-05
Reason
Incorrect IOL diopter
Root cause
Process control
Action
On 04/17/2020, the firm sent an "Urgent: Voluntary Medical Device Removal' letter to customers via overnight mail that the firm became aware that there was a potential problem relating to the IOL of wrong sized diopter, in a pre-loaded delivery system. Should a patient be implanted with an IOL having incorrect optical power, a refractive error would result. Patients with a refractive error may experience symptoms from mild blurred vision to decreased vision, which may require additional interventions such as spectacles or surgical intervention to correct. The firm is instructing customers to: (i) Return any unused product of specified lot number (ii) The firm's Customer Service will contact customers to arrange for the return and replacement of affected Pre-loaded delivery system (iii) Fill out "Response Form" (iv) Return the "Response form" via fax or email to the firm, provided of contact information: Fax No.: 817-302-4337; E-mail: Market.Actions@alcon.com Additionally, the firm is asking customers to: Reach out to firm's Sales Representative for any questions.
Quantity
84 units
Distribution
US Nationwide distribution including in the states of TX, PA, CT, IA, MI, NE.
Lot, serial or model codes
Lot number 12726594; Model Number: AU00T0V205; UDI # 038065GMN000065H7; 0380652394772(17)220630(21)12726594000(30)1
510(k) numbers
not on file
PMA numbers
["P930014S084"]
Product code
HQL
Firm FEI
1610287
Firm city
Fort Worth
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register