Skip to the content

FDA Medical Device Recalls (openFDA)

Luminex Corporation: ARIES HSV 1&2 Assay , REF 50-10017, UDI # 00840487100295

Recall number
Z-2283-2019
Recalling firm
Luminex Corporation
Product
ARIES HSV 1&2 Assay , REF 50-10017, UDI # 00840487100295
Status
Open, Classified
Initiated
2019-02-20
Posted
2019-08-15
Terminated
not on file
Reason
Through internal investigation, it was discovered that a portion of side cassette assemblies (SCAs) back side seal (BSS) did not meet design specification, resulting in partially sealed SCAs. The impact of a partial seal of the BSS could result in potential decrease of volume for each solution in the SCAs and decreased sensitivity or stability of the assay.
Root cause
Process control
Action
On February 20, 2019, the firm sent a "Urgent: Medical Device Recall/Field Safety Notice" via email to all consignees. The customer notification informed customers of the following: 1. As an immediate action, Luminex is requesting that customers discontinue use of and discard the products listedin the recall notice immediately or by the Discontinue Use By date. Contact Luminex Global Support Services to obtain replacement material. Material that is still within its Discontinue Use By date may be used at the customers discretion. 2. Complete the Acknowledgment and Receipt form. You must complete and return this form even if you do not have any remaining product. Luminex Technical Support can assist you in completing this form. This information is essential in order to maintain recall effectiveness information required by US FDA. 3. Although no adverse events have been reported, adverse reaction or quality problems experienced with the use of this product may be reported to the FDAs MedWatch Adverse Event Reporting program either online, by regular mail or by fax. 4. Please call the Luminex Global Support Services Team if you have any questions or concerns. Luminex Global Support Services 1-877-785-2323 (U.S. and Canada) +1-512-381-4397 (Outside U.S. and Canada) support@luminexcorp.com CAN-0233
Quantity
not on file
Distribution
US: AK, AL, AZ,CA,CO,FL,GA,IL,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,NE,NJ,NM,NY,OH,ON,OR,PA,TN,TX,UT,VA,VT,WA,WI, OUS: Belgium, Canada, Thailand, Germany, Singapore, United Kingdom
Lot, serial or model codes
Lot numbers & discontinue use by dates for Part Number 50-10017:  AA7308 - Immediately; AA7485 - 2/23/2019; AA7545 - 3/13/2019; AA7625 - 4/10/2019; AA7630 - 3/16/2019; AA7636 - 4/3/2019; AA7945 - 4/25/2019; AA7985 - 5/8/2019
510(k) numbers
["K151906","K161495"]
PMA numbers
not on file
Product code
PGI
Firm FEI
3002524000
Firm city
Austin
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register