FDA Medical Device Recalls (openFDA)

Covidien, LLC: Covidien Cytosponge Cell Collection Device, REF: CYTO-201; and Covidien Cytosponge Cell Collection Kit, REF: CYTO-KIT-USM

Recall number
Z-2282-2023
Recalling firm
Covidien, LLC
Product
Covidien Cytosponge Cell Collection Device, REF: CYTO-201; and Covidien Cytosponge Cell Collection Kit, REF: CYTO-KIT-USM
Status
Open, Classified
Initiated
2023-06-02
Posted
2023-08-02
Terminated
not on file
Reason
Cell Collection Device may be at increased risk of the sponge detaching from the string during removal of the device from the patient, which could lead to device fragments in patient, obstruction, airway obstruction, secondary intervention, secondary intervention (with the primary procedure), supraglottic airway obstruction, and aspiration.
Root cause
Process control
Action
On 6/2/23, Medtronic sent recall notices to customers who were informed of the following: 1) If product was purchased directly from Medtronic, return affected product to Medtronic. 2) If product was purchased from a distributor, return affected product to the distributor. 3) The recall notice should be passed on to all who need to be aware within your organization or to any organization where the potentially affected product has been transferred. 4) Complete and return the Customer Confirmation Form to rs.gmbfcamitg@medtronic.com Adverse reactions or quality problems experienced with this product can be reported to Technical Support at dl.gi-usdl-ustecchsupport@medtronic.com or call 1-800-448-3644 option 3.
Quantity
6902
Distribution
Worldwide - US Nationwide distribution in the states of NY, MN, VA, WI and the countries of Canada, Netherlands, Poland, United Kingdom.
Lot, serial or model codes
REF/UDI-DI/Lot: CYTO-201/20884521767789/22E1619JZX, 22F0680JZX, 22F0950JZX, 22G0007JZX, 22G0008JZX, 22G0723JZX, 22G0724JZX, 22H0017JZX, 22H0018JZX, 22H0019JZX, 22H0820JZX, 22H0821JZX, 22H1158JZX, 22H1159JZX, 22H1161JZX, 22J0162JZX, 22J0163JZX, 22J0164JZX, 22J1030JZX, 22K0966JZX, 22K0967JZX, 23B0301JZX, 23B0303JZX; CYTO-KIT-USM/10884521546257/22G0635JZX, 22E1624JZX, 22E1625JZX.
510(k) numbers
["K181020"]
PMA numbers
not on file
Product code
EOX
Firm FEI
3004904811
Firm city
Santa Clara
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register