Konica Minolta Medical Imaging USA, Inc.: ImagePilot, Model No. D9MA

Recall number
Z-2282-2018
Recalling firm
Konica Minolta Medical Imaging USA, Inc.
Product
ImagePilot, Model No. D9MA
Status
Terminated
Initiated
2018-05-09
Posted
not on file
Terminated
2020-04-10
Reason
Panasonic has issued a Mandatory Product Safety Notice, indicating a possibility of battery ignition (fire) in CF-C2 "Toughbook" laptops. Since Konica Minolta uses this laptop in some of its product configurations, it notified its customers to utilize a BIOS Utility for the laptops to minimize the risk of an incident while Panasonic develops a final solution.
Root cause
Unknown/Undetermined by firm
Action
The firm issued an Important Product Recall Information letter to their affected customers dated May 3, 2018, delivered via UPS Overnight Delivery. The letter notified customers of the Panasonic recall and how it affected the Konica Minolta product. The letter instructed customers to utilize a BIOS Utility for CF-C2 laptops to minimize the risk of an incident while Panasonic develops a final solution. Customers were provided with Technical Bulletin which detailed the steps. Correction Option 1: Update BIOS Environmental Setting using Utility Tool 1. Download the BIOS Utility Tool from the Panasonic Website: ftp://ftp.panasonic.com/ computer/ software/ Panasonic-CF-C2-Charge-Tool.zip 2. Load the Utility Tool onto the CF-C2 Desktop 3. Double-click to run batch file: ChgCtrlUty.bat 4. Optional tool: If Integrated Portable, an external keyboard may be required. Visit KM University at https://km-university.com/ for video instructions: https://km-university.com/course/notification-of-mandatory-safety-bulletin-of-panasonic-toughbook-cf-c2/ Correction Option 2: Manual change to the BIOS Environmental Setting (for Integrated Portable, an external keyboard will be required) 1. If power is on, turn the power off and restart prior to following instructions. 2. Upon boot-up, at Panasonic splash screen, continuously hit the F2 key on the keyboard to access the BIOS. a. Go to the "MAIN" tab associated with the BIOS setting. b. Select "ENVIRONMENT Current Status" setting. c. Ensure "HIGH TEMPERATURE" is selected. Otherwise change this setting to high and click ENTER. d. Go to the "EXlr' tab associated with the BIOS setting. e. Select "SAVE VALUES AND REBOOr' and hit ENTER key on the keyboard. f. A prompt will appear to state, "SAVE VALUES". Select YES and hit ENTER. If you have any questions, need a keyboard, require assistance with these instructions, and/or a safety verification please contact our Customer Care Team at (800) 945-045
Quantity
not on file
Distribution
Nationwide distribution. Foreign distribution to Argentina, Brazil, Canada, Chile, Colombia, Ecuador, Japan, Mexico, Netherlands, and Peru.
Lot, serial or model codes
All version numbers.
510(k) numbers
["K071436"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
2241281
Firm city
Wayne
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register