FDA Medical Device Recalls (openFDA)

Stryker Howmedica Osteonics Corp.: Stryker Orthopaedics Trident Constrained Liner Inserter/Impactor Tip, 22 mm, 28 mm, and 32 mm. Non Sterile Catalog Numbers 2199-2022, 2199-2028 and 2199-2032. For use in total hip arthroplasty.

Recall number
Z-2281-2016
Recalling firm
Stryker Howmedica Osteonics Corp.
Product
Stryker Orthopaedics Trident Constrained Liner Inserter/Impactor Tip, 22 mm, 28 mm, and 32 mm. Non Sterile Catalog Numbers 2199-2022, 2199-2028 and 2199-2032. For use in total hip arthroplasty.
Status
Terminated
Initiated
2016-06-03
Posted
not on file
Terminated
2018-08-01
Reason
Impactors do not meet the required sterility assurance level (SAL) of 10-6 when using the recommended steam sterilization method.
Root cause
Process design
Action
Stryker notified its branches/agencies of this recall by e-mail on May 26, 2016. They were asked to quarantine the affected devices. Stryker sent a recall letter and business reply form dated June 3, 2016 to their affected customers via UPS. Stericycle will be handling the returns.
Quantity
1,864 units
Distribution
Worldwide distribution. US nationwide (AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MS, NJ, NY, OH, PA, TX, WA, WI, WV), Sweden, South Africa, Chile, Australia, New Zealand, Canada, United Kingdom, Switzerland, Colombia, and Poland.
Lot, serial or model codes
Catalog Nos. 2199-2022, 2199-2028 and 2199-2032 All lots manufactured from May 30, 2012
510(k) numbers
["K151264"]
PMA numbers
not on file
Product code
LPH
Firm FEI
3003070421
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register