Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Corporation: Stryker SmartPump Tourniquet, disposable sterile (single use) DISP 18X3,1BLA,1PRT QUICK / 5921-018-135; DISP 18X3,1BLA,2PRT QUICK / 5921-018-235; DISP 24X4,1BLA, 2PRT QUICK / 5921-024-235

Recall number
Z-2277-2025
Recalling firm
Stryker Corporation
Product
Stryker SmartPump Tourniquet, disposable sterile (single use) DISP 18X3,1BLA,1PRT QUICK / 5921-018-135; DISP 18X3,1BLA,2PRT QUICK / 5921-018-235; DISP 24X4,1BLA, 2PRT QUICK / 5921-024-235
Status
Open, Classified
Initiated
2025-07-02
Posted
2025-08-07
Terminated
not on file
Reason
Tourniquet cuff flange may become detached from the bladder. If leak occurs during the procedure, there may be risk of hemorrhage, resulting in blood loss greater than expected. Other serious risks include cerebral hypotension and cardiovascular collapse and arrest, risking stroke or death. Specific to IVRA, patient toxicity may also occur due to rapid systemic absorption of anesthetics.
Root cause
Equipment maintenance
Action
On July 2, 2025, the firm notified customers via Urgent Medical Device Recall letters. Customers were instructed to locate and quarantine any affected products at their facility and return an enclosed business reply form to the firm. Stryker will provide a shipping label to return the affected product, and credit will be issued to affected customers upon receipt of the recalled product.
Quantity
233.230
Distribution
Domestic US distribution nationwide. International distribution to Australia Canada Hong Kong India Malaysia Netherlands New Zealand Singapore Sweden Thailand United Kingdom
Lot, serial or model codes
5921-018-135 GTIN 07613154599257 5921-018-235 GTIN 07613154599295 5921-024-235 GTIN 07613154599370
510(k) numbers
not on file
PMA numbers
not on file
Product code
KCY
Firm FEI
3015967359
Firm city
Portage
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register