Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: SOMATOM Force, Model Number 10742326 - Product Usage: Computed tomography systems intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy planning. These CT systems can be used for low dose lunch cancer screening in high risk populations (as defined by professional medical societies).

Recall number
Z-2274-2019
Recalling firm
Siemens Medical Solutions USA, Inc
Product
SOMATOM Force, Model Number 10742326 - Product Usage: Computed tomography systems intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy planning. These CT systems can be used for low dose lunch cancer screening in high risk populations (as defined by professional medical societies).
Status
Terminated
Initiated
2019-07-03
Posted
not on file
Terminated
2024-05-10
Reason
Potential for four safety-related software issues impacting systems utilizing syngo.CT software version VB10A which may result in a scan abort.
Root cause
Software design
Action
Siemens has initiated a Customer Safety Advisory Notice (CSAN) to all affected customers via CT035/19/S. Customers with valid email addresses were emailed a copy of the CSAN on 07/03/2019 via Adobe Mega Sign. The CSAN were distributed to the remaining customers without a valid email address via certified mail on 07/12/2019. Additionally, a solution to eliminate the root cause of this problem (application of software update VB20) will be distributed to all affected customers free of charge. Following the corrective action, the cause is eliminated and a recurrence of the fault is prevented.
Quantity
not on file
Distribution
US Nationwide distribution.
Lot, serial or model codes
Serial Numbers: 75765 75949 94851 75984 76019 75789 75512 75770 75600 75994 76046 75768 75974 75867 75510 75535 75820 75754 75755 75993 75916 75971 75893 75944 75540 76027 75980 75571 75848 75953 75885 75500 75973 75850 75527 75728 75939 75923 83819 76057 75450 75760 75806 75808 75476 75588 75931 75900 75981 75906 75602 75964 75804 75837 75787 75743 76010 76041 75936 75959 75905 75942 75941 75805 75884 76006 75919 76003 75913 76050 76016 75524 75778 75501 75570 75769 75983 75849 76000 75999 75976 76008 75616 75750 76001
510(k) numbers
["K133589"]
PMA numbers
not on file
Product code
JAK
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register