Beckman Coulter Inc.: Allegra X-5 rotor yoke, Catalog No. (Instruments) B30589 and B30590 (Rotor) B30593. Product Labeling: IFU B29071AA. Product Usage: For In Vitro Diagnostic Use. The Allegra X-5 is intended for the separation of components through the use of relative centrifugal force. It is designed to separate human samples, including blood and other body fluids, for processing, analysis and in vitro diagnostic testing, as well as non-human body samples and chemicals, including industrial and environmental samples. This centrifuge should be operated by qualified personnel only.

Recall number
Z-2274-2016
Recalling firm
Beckman Coulter Inc.
Product
Allegra X-5 rotor yoke, Catalog No. (Instruments) B30589 and B30590 (Rotor) B30593. Product Labeling: IFU B29071AA. Product Usage: For In Vitro Diagnostic Use. The Allegra X-5 is intended for the separation of components through the use of relative centrifugal force. It is designed to separate human samples, including blood and other body fluids, for processing, analysis and in vitro diagnostic testing, as well as non-human body samples and chemicals, including industrial and environmental samples. This centrifuge should be operated by qualified personnel only.
Status
Terminated
Initiated
2016-02-05
Posted
not on file
Terminated
2017-01-17
Reason
A defective rotor yoke on Allegra X-5 allows improper rotor seating on the hub resulting in possible rotor failure if the buckets are loaded in an imbalanced configuration.
Root cause
Nonconforming Material/Component
Action
Beckman Coulter sent an Important Product Notice Recall letter dated February 5, 2016 to customer. The letter identified the affected product problem and actions to be taken. Customers are instructed to complete and return the enclosed Response Form. Customers with questions are instructed to contact our Customer Support Center " Via website, http://www.beckmancoulter.com/customersupport/support " Via phone, call 800-369-0333 the United States and Canada " Outside the United States and Canada, contact your local Beckman Coulter Representative.
Quantity
62 units total (33 units in US)
Distribution
Worldwide Distribution - US Nationwide (including Puerto Rico) and the countries of: Australia, Canada, Colombia, Germany, Guatemala, Netherlands, Singapore, Tunisia, , Spain
Lot, serial or model codes
AFB13H005, AFB13H008, AFB13H012, AFB13K002, AFB13K005, AFB13K008, AFB13J002, AFB13J003, AFB13J004, AFB13J005, AFB13J007, AFB15J001, AFB15C010, AFB13K009, AFB14H001, AFB15C008, AFB15C003, AFB15C005, AFB15C004, AFB15C006, AFB14H005, AFB14H003, AFB13K004, AFB13K006, AFB13K007, AFB13J008, AFB13K001, AFB13J006, AFB13J010, AFB13H013, AFB13H009, AFB13H003, AFB13H004
510(k) numbers
not on file
PMA numbers
not on file
Product code
JQC
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register