FDA Medical Device Recalls (openFDA)
Alphatec Spine, Inc.: TRESTLE LUXE Anterior Cervical Plating System, Product code KWQ. 510(k) K102820. The TRESTLE LUXE Anterior Cervical Plating System is a temporary device used to stabilize the cervical spine during bone fusion development. Device implants include a range of plate sizes and bone screws to provide the versatility required for the specific indications noted. Fixation is achieved by means of a rigid plate that is surgically attached to the spine with bone screws. The TRESTLE LUXE Anterior Cervical Plating System is intended for use in the anterior cervical spine (C2-C7). Patients with trauma (including fractures), spondylolisthesis, and pseudoarthrosis, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spinal stenosis, tumors, and failed previous fusion.
- Recall number
- Z-2273-2012
- Recalling firm
- Alphatec Spine, Inc.
- Product
- TRESTLE LUXE Anterior Cervical Plating System, Product code KWQ. 510(k) K102820. The TRESTLE LUXE Anterior Cervical Plating System is a temporary device used to stabilize the cervical spine during bone fusion development. Device implants include a range of plate sizes and bone screws to provide the versatility required for the specific indications noted. Fixation is achieved by means of a rigid plate that is surgically attached to the spine with bone screws. The TRESTLE LUXE Anterior Cervical Plating System is intended for use in the anterior cervical spine (C2-C7). Patients with trauma (including fractures), spondylolisthesis, and pseudoarthrosis, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spinal stenosis, tumors, and failed previous fusion.
- Status
- Terminated
- Initiated
- 2012-05-29
- Posted
- 2012-08-27
- Terminated
- 2012-09-04
- Reason
- On May 9th 2012 Alphatec Spine received a complaint which reported that the distal tip of the Trestle Luxe Slide Alignment Tool (Part Number 71721), does not properly fit into the hexalobe screw on the Trestle Luxe Plate.
- Root cause
- Device Design
- Action
- Alphatec Spine sent an "URGENT-MEDICAL DEVICE RECALL NOTIFICATION" letter dated May 29, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. A form was attached for customers to complete and return via fax to 760-431-0289. Contact Alphatec Spine's Customer Service Department at customerservice@alphatecspine.comfor questions regarding this recall.)
- Quantity
- 38
- Distribution
- Nationwide Distribution.
- Lot, serial or model codes
- Product code KWQ. 510(k) K102820 Part Number 71721. Lot numbers: 6629701, 6594701, 6434301, 6434302, 6434302R1.
- 510(k) numbers
- ["K102820"]
- PMA numbers
- not on file
- Product code
- KWQ
- Firm FEI
- 2027467
- Firm city
- Carlsbad
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28