Beckman Coulter Inc.: UniCel DxC Synchron Clinical Systems and UniCel DxC Synchron Access Integrated Clinical Systems The affected part is Part number 70300. Part number 970300 is used to transfer No Foam Solution from the No Foam Bottle to the Waste B canister and the Waste Manifold on the DxC Hydro System. Catalog No. UniCel DxC 800 Synchron Clinical Systems UniCel DxC 600 Synchron Clinical Systems UniCel DxC 600i Synchron Access Integrated Clinical Systems UniCel DxC 880i Synchron Access Integrated Clinical Systems UniCel DxC 660i Synchron Access Integrated Clinical Systems Product Labeling Reference No. A93719 For the in vitro determination of a variety of general chemistries, therapeutic drugs, and other chemistries.
- Recall number
- Z-2272-2016
- Recalling firm
- Beckman Coulter Inc.
- Product
- UniCel DxC Synchron Clinical Systems and UniCel DxC Synchron Access Integrated Clinical Systems The affected part is Part number 70300. Part number 970300 is used to transfer No Foam Solution from the No Foam Bottle to the Waste B canister and the Waste Manifold on the DxC Hydro System. Catalog No. UniCel DxC 800 Synchron Clinical Systems UniCel DxC 600 Synchron Clinical Systems UniCel DxC 600i Synchron Access Integrated Clinical Systems UniCel DxC 880i Synchron Access Integrated Clinical Systems UniCel DxC 660i Synchron Access Integrated Clinical Systems Product Labeling Reference No. A93719 For the in vitro determination of a variety of general chemistries, therapeutic drugs, and other chemistries.
- Status
- Terminated
- Initiated
- 2014-04-22
- Posted
- 2016-07-22
- Terminated
- 2017-02-09
- Reason
- Possible leaking of No Foam solution from the No Foam Bottle Assembly on the UniCel DxC Synchron Systems.
- Root cause
- Nonconforming Material/Component
- Action
- An Important Product Notice dated 4/22/14 was sent to all their customers to inform them that Beckman Coulter has identified a manufacturing issue with a plastic fitting on the No Foam Bottle Assembly which may crack and cause No Foam solution to leak. The letter informs the customers of the impact, action to be taken, and resolution. Customers are asked to inspect the No Foam Bottle Assembly for leaks prior to refilling the bottle. If leaks occur, customers should contact Customer Support Center following the instructions provided. Customers are instructed to complete the enclosed response form within 10 days. Customers with questions are instructed to contact the firm's Customer Support Center - Via our website, http://www.beckmancoulter.com/customersupport/support - Via phone, call 1-800-854-3633 in the United States and Canada - Outside the United States and Canada, contact your local Beckman Coulter Representative.
- Quantity
- 3,540 units total (1,067 units in US)
- Distribution
- Worldwide Distribution -- United Arab Emirates, Angola, Australia, Belgium, Bulgaria, Canada, Switzerland, China, Colombia, Costa Rica , Czech Republic, Germany, Denmark, Egypt, Eritrea, Spain, Finland, France, United Kingdom, Greece, Guatemala, Hong Kong, Indonesia, Ireland, India, Italy, Jordan, Japan, Korea, Republic of, Kuwait, Lebanon, Lithuania, Morocco, Mexico, Malaysia, Namibia, Netherlands, Norway, New Zealand, Oman, Panama, Philippines, Pakistan, Poland, Portugal, Paraguay, Qatar, Romania, Serbia, Russian Federation, Rwanda, Saudi Arabia, Sweden, Singapore, Slovakia, Swaziland, Thailand, Tunisia, Turkey, Taiwan, US (including Puerto Rico), Vietnam, South Africa, and Zimbabwe.
- Lot, serial or model codes
- Up to and including 6133
- 510(k) numbers
- ["K042291","K060256","K103842"]
- PMA numbers
- not on file
- Product code
- JJE
- Firm FEI
- 2050012
- Firm city
- Brea
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28