FDA Medical Device Recalls (openFDA)
Invacare Corporation: Invacare Power Wheelchair, one per box To provide mobility to a person restricted to a sitting position.
- Recall number
- Z-2270-2013
- Recalling firm
- Invacare Corporation
- Product
- Invacare Power Wheelchair, one per box To provide mobility to a person restricted to a sitting position.
- Status
- Terminated
- Initiated
- 2013-06-04
- Posted
- 2013-09-23
- Terminated
- 2016-01-04
- Reason
- Unintended slowing or deceleration, and on rare occasion, unanticipated and unattended acceleration of power wheelchairs equipped with SPJ and MPJ type joysticks.
- Root cause
- Process control
- Action
- Invacare sent an Urgent Patient Safety Notification letter to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to contact their wheelchair povider to have their joystick replaced. Customers with questions were instructed to call their provider at Invacare's recall support line at 877-413-6008. Go to www.invacare.com/joystickrecall for more information regarding this recall
- Quantity
- 166,230
- Distribution
- Worldwide Distribution - USA including AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. Internationally to Canada, Australia, Bermuda, Brazil, China, Costa Rica, Germany, France, United Kingdom, Guam, Hong Kong, Israel, India, Kuwait, New Zealand, Puerto rico, Qatar, Saudi Arabia, Singapore, Thailand, Trinidad, Tobago, and South Africa.
- Lot, serial or model codes
- The following Joystick model numbers and respective part numbers are affected by this recall: CSPJ, 1121239; SPJPLUS, 1127291; SPJAP, 1136937; SPJAPM, 1136938; CMPJM6, 1164361; CMPM6U, 1644361; MPJM6, 1136885; MPJM6U, 1136885; 1558M6, 1136887; 1558MM6, 1136887; 1500M6, 1136889; PACM6, 1136888; PACM6, 1167927; PSRM6, 1139941; PSRM6, 1140046; PSRM6, 1140047; PSRM6, 1140048; PSRM6, 1144844; PSRM6, 1144845; and PSFM6, 1139944. The following Power Wheelchair model numbers (affected date ranges are included in the 1525712 6-3-13 001C model numbers FINAL xlsx document) are affected by this recall: M61UKRD; 3GAR; 3GARBASE; 3GAR-CG; 3GRX; 3GRXBASE; 3GRX-CG; 3GSEAT-CG; 3GSTRX-TS; 3GTQ3; 3GTQ3-CG; 3GTQ3-MCG; 3GTQ3V; 3GTQ-CG; 3GTQ-MCG; 3GTQSE; 3GTQSEBASE; 3GTQSE-CG; 3GTQSE-PTO; 3GTQSP; 3GTQSPBASE; 3GTQSP-TS; FDX; FDX-CG; FDX-MCG; M41BB; M41RB; M41RBCN; M51 PBLUEBASE; M51 PREDBASE; M51BB; M51-CG; M51CGBB; M51CGRB; M51P; M51PBB; M51PBLUEBASE; M51PR; M51PRB; M51PREDBASE; M51RB; M51RBCN; M51SEAT-CG; M61; M61 PBLUE; M61 PRED ; M61 ; KRDBASE; M61(1144227); M61FBLUE; M61FRAD; M61FRED; M61R; M61RB(1144227); M61RBCN; M61RED; M61SEAT; M61UK; M61UK REDBASE; M61UKRD; M61UKRDBASE; M71; M71JR; M71R; M71SEAT; M71-TS; M91; M91-C; M91-M; M91R; M91-R; M91SEAT; M91-TS; M94; M94-C; PTBASE; PTORB; PWORBITS; R51; R51LXP; TDX3; TDX3BASE; TDX3-PS; TDX3SE; TDX3SEBASE; TDX3SE-PS; TDX3SE-TS; TDX3-TS; TDX4; TDX4BASE; TDX4-PS; TDX4-TS; TDX5; TDX5BASE; TDX5-PS; TDX5-TS; TDXSC; TDXSC2-CG; TDXSCSEAT; TDXSCV; TDXSEAT; TDXSI; TDXSI-2; TDXSI-2-S; TDXSI-CG; TDXSI-HD; TDXSI-HD-S; TDXSISEAT; TDXSISEAT-HD; TDXSIV; TDXSIV-2; TDXSIV-2-S; TDXSIV-HD; TDXSIV-HD-S; TDXSP; TDXSPBASE; TDXSP-CG; TDXSP-CG-GT; TDXSP-GT; TDXSP-MCG; TDXSP-MCG-GT; TDXSPREE; TDXSPREE-CG; TDXSPREESEAT; TDXSPSEAT; TDXSR; TDXSR-CG; TDXSR-CG-HD; TDXSR-HD; TDXSR-MCG; TDXSR-MCG-HD; TDXSRSEAT; TDXSRV; and TDXSRV-HD.
- 510(k) numbers
- ["K023589","K993413"]
- PMA numbers
- not on file
- Product code
- ITI
- Firm FEI
- 1525712
- Firm city
- Elyria
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as