Skip to the content

FDA Medical Device Recalls (openFDA)

Invacare Corporation: Invacare Power Wheelchair, one per box To provide mobility to a person restricted to a sitting position.

Recall number
Z-2270-2013
Recalling firm
Invacare Corporation
Product
Invacare Power Wheelchair, one per box To provide mobility to a person restricted to a sitting position.
Status
Terminated
Initiated
2013-06-04
Posted
2013-09-23
Terminated
2016-01-04
Reason
Unintended slowing or deceleration, and on rare occasion, unanticipated and unattended acceleration of power wheelchairs equipped with SPJ and MPJ type joysticks.
Root cause
Process control
Action
Invacare sent an Urgent Patient Safety Notification letter to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to contact their wheelchair povider to have their joystick replaced. Customers with questions were instructed to call their provider at Invacare's recall support line at 877-413-6008. Go to www.invacare.com/joystickrecall for more information regarding this recall
Quantity
166,230
Distribution
Worldwide Distribution - USA including AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. Internationally to Canada, Australia, Bermuda, Brazil, China, Costa Rica, Germany, France, United Kingdom, Guam, Hong Kong, Israel, India, Kuwait, New Zealand, Puerto rico, Qatar, Saudi Arabia, Singapore, Thailand, Trinidad, Tobago, and South Africa.
Lot, serial or model codes
The following Joystick model numbers and respective part numbers are affected by this recall:  CSPJ, 1121239; SPJPLUS, 1127291; SPJAP, 1136937; SPJAPM, 1136938; CMPJM6, 1164361; CMPM6U, 1644361; MPJM6, 1136885; MPJM6U, 1136885; 1558M6, 1136887; 1558MM6, 1136887; 1500M6, 1136889; PACM6, 1136888; PACM6, 1167927; PSRM6, 1139941; PSRM6, 1140046; PSRM6, 1140047; PSRM6, 1140048; PSRM6, 1144844; PSRM6, 1144845; and PSFM6, 1139944.   The following Power Wheelchair model numbers (affected date ranges are included in the 1525712 6-3-13 001C model numbers FINAL xlsx document) are affected by this recall: M61UKRD; 3GAR; 3GARBASE; 3GAR-CG; 3GRX; 3GRXBASE; 3GRX-CG; 3GSEAT-CG; 3GSTRX-TS; 3GTQ3; 3GTQ3-CG; 3GTQ3-MCG; 3GTQ3V; 3GTQ-CG; 3GTQ-MCG; 3GTQSE; 3GTQSEBASE; 3GTQSE-CG; 3GTQSE-PTO; 3GTQSP; 3GTQSPBASE; 3GTQSP-TS; FDX; FDX-CG; FDX-MCG; M41BB; M41RB; M41RBCN; M51 PBLUEBASE; M51 PREDBASE; M51BB; M51-CG; M51CGBB; M51CGRB; M51P; M51PBB; M51PBLUEBASE; M51PR; M51PRB; M51PREDBASE; M51RB; M51RBCN; M51SEAT-CG; M61; M61 PBLUE; M61 PRED ; M61 ; KRDBASE; M61(1144227); M61FBLUE; M61FRAD; M61FRED; M61R; M61RB(1144227); M61RBCN; M61RED; M61SEAT; M61UK; M61UK REDBASE; M61UKRD; M61UKRDBASE; M71; M71JR; M71R; M71SEAT; M71-TS; M91; M91-C; M91-M; M91R; M91-R; M91SEAT; M91-TS; M94; M94-C; PTBASE; PTORB; PWORBITS; R51; R51LXP; TDX3; TDX3BASE; TDX3-PS; TDX3SE; TDX3SEBASE; TDX3SE-PS; TDX3SE-TS; TDX3-TS; TDX4; TDX4BASE; TDX4-PS; TDX4-TS; TDX5; TDX5BASE; TDX5-PS; TDX5-TS; TDXSC; TDXSC2-CG; TDXSCSEAT; TDXSCV; TDXSEAT; TDXSI; TDXSI-2; TDXSI-2-S; TDXSI-CG; TDXSI-HD; TDXSI-HD-S; TDXSISEAT; TDXSISEAT-HD; TDXSIV; TDXSIV-2; TDXSIV-2-S; TDXSIV-HD; TDXSIV-HD-S; TDXSP; TDXSPBASE; TDXSP-CG; TDXSP-CG-GT; TDXSP-GT; TDXSP-MCG; TDXSP-MCG-GT; TDXSPREE; TDXSPREE-CG; TDXSPREESEAT; TDXSPSEAT; TDXSR; TDXSR-CG; TDXSR-CG-HD; TDXSR-HD; TDXSR-MCG; TDXSR-MCG-HD; TDXSRSEAT; TDXSRV; and TDXSRV-HD.
510(k) numbers
["K023589","K993413"]
PMA numbers
not on file
Product code
ITI
Firm FEI
1525712
Firm city
Elyria
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.