FDA Medical Device Recalls (openFDA)
Smiths Medical ASD, Inc.: Smiths Medical, Deltec: REF 21-8011-24 Port-A-Cath Dual-Lumen. Titanium Venous Access System. REF 21-8052-24 & REF 21-8053-24 , Port-A-Cath II Dual-Lumen, Polysulfone /Titanium Venous Access System. Sterile EO, Rx Only. PORT-A-CATH and PORT-A-CATH II implantable venous and arterial access systems are designed to permit repeated access to the vascular system for the parenteral delivery of medications, fluids, and nutritional solutions and for the sampling of venous blood. Introducer sets are indicated for the introduction of catheters into the subclavian vein.
- Recall number
- Z-2269-2013
- Recalling firm
- Smiths Medical ASD, Inc.
- Product
- Smiths Medical, Deltec: REF 21-8011-24 Port-A-Cath Dual-Lumen. Titanium Venous Access System. REF 21-8052-24 & REF 21-8053-24 , Port-A-Cath II Dual-Lumen, Polysulfone /Titanium Venous Access System. Sterile EO, Rx Only. PORT-A-CATH and PORT-A-CATH II implantable venous and arterial access systems are designed to permit repeated access to the vascular system for the parenteral delivery of medications, fluids, and nutritional solutions and for the sampling of venous blood. Introducer sets are indicated for the introduction of catheters into the subclavian vein.
- Status
- Terminated
- Initiated
- 2013-09-10
- Posted
- 2013-09-23
- Terminated
- 2014-10-21
- Reason
- Smiths Medical is conducting a voluntary recall of PORT-A-CATH and PORT-A-CATH II Implantable Venous and Arterial Access Systems and Introducer Sets. Smiths Medical has become aware of a small number of customer reports regarding the introducer sheath and the introducer opening being too small for the width of the catheter with the use of these Sets. Smiths Medical has received no reports
- Root cause
- Mixed-up of materials/components
- Action
- Smiths Medical ASD, Inc. sent an Urgent Field Safety Notice dated September 6, 2013, to all affected customers. The letter described the problem and the product involved in the recall. The letter included the "Advice on Action to be Taken by the User" which instructed consignees to inspect and remove affected devices and to complete and return the "Urgent Field Safety Notice Confirmation Form. The letter also included information on the "Transmission of this Urgent Field safety Notice" which included information where customers should report issues related to the product to Smiths Medical's Global Complaint Department at 1-866-216-8806 or globalcomplaints@smiths-medical.com. For questions on the information provided they should contact 800-258-5361 (Smiths Medical's Customer Service Department). For questions regarding this recall call 1-800-332-0178.
- Quantity
- 372
- Distribution
- Worldwide Distribution - USA including AZ, AR, CA, GA, IL, IN, KY, MD, MA, MI, NV, NJ, NY, NC, OH, OR, PA, RI, SC, TN, TX, VA, WA, WV, WY. Internationally to DENMARK, BELGIUM CANADA.
- Lot, serial or model codes
- 21-8011-24: 1971567 1979805 1984078 1997124 2006124 2038001 2090898 2151160 2196530 2216256 2249695 21-8052-24: 1971572 2012472 2022904 2067947 2080704 2146472 2168988 2196532 2207404 2237458 2249696 2269178 2282158 21-8053-24: 2168989 2249697
- 510(k) numbers
- ["K932095","K932841","K955407"]
- PMA numbers
- not on file
- Product code
- LJT
- Firm FEI
- 2183502
- Firm city
- Saint Paul
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as