FDA Medical Device Recalls (openFDA)
Alere San Diego, Inc.: Henry Schein One Step + Mono Rapid Test Device, Catalog No. 9004070 a simple test that utilizes an extract of bovine erythrocytes to qualitatively and selectively detect IM heterophile antibodies in whole blood, serum or plasma in just minutes.
- Recall number
- Z-2268-2016
- Recalling firm
- Alere San Diego, Inc.
- Product
- Henry Schein One Step + Mono Rapid Test Device, Catalog No. 9004070 a simple test that utilizes an extract of bovine erythrocytes to qualitatively and selectively detect IM heterophile antibodies in whole blood, serum or plasma in just minutes.
- Status
- Terminated
- Initiated
- 2016-06-22
- Posted
- not on file
- Terminated
- 2018-01-29
- Reason
- The Positive Control information referencing human plasma as the antibody material described in the product labeling and packaging is incorrect.
- Root cause
- Unknown/Undetermined by firm
- Action
- A Medical Device Correction letter dated 6/22/16 will be sent to all customers to inform them that Alere is recalling the MONO Rapid Test products because the Positive Control Information referencing human plasma as the antibody material described in the product labeling and packaging is incorrect. Customers are informed that the labeling states the Positive Control contains the following: Diluted human plasma containing IM heterophile antibodies, 0.09% sodium azide. The Positive Control contains the following: Goat anti-mono antibody, 0.09% sodium azide. Customers are instructed to complete and fax or email the enclosed verification form within 10 days. Customers with questions are instructed to contact Alere San Diego at (844) 804-4945 or email at verfications.ts@alere.com.
- Quantity
- not on file
- Distribution
- US
- Lot, serial or model codes
- all lots are affected
- 510(k) numbers
- ["K042272"]
- PMA numbers
- not on file
- Product code
- KTN
- Firm FEI
- 1000125596
- Firm city
- San Diego
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as