Beckman Coulter, Inc.: DxI 800 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71457

Recall number
Z-2267-2025
Recalling firm
Beckman Coulter, Inc.
Product
DxI 800 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71457
Status
Open, Classified
Initiated
2025-06-25
Posted
2025-08-05
Terminated
not on file
Reason
Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.
Root cause
Device Design
Action
An URGENT MEDICAL DEVICE RECALL notification dated 6/25/25 was sent to customers. ACTION: If your laboratory encounters any of the events described in this letter: " Perform a system initialization to resume normal operation. o If repeated initialization attempts fail, contact Customer Technical Support. " Repeat testing for any cancelled tests. " No action is required if your laboratory has not observed the events listed above ESOLUTION: "A Beckman Coulter representative will schedule the modification installation when it is available. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product listed above to another laboratory, please provide them a copy of this letter. So that we are assured you have received this important communication, please respond within 10 days in one of the following ways: " Electronically, if you received this communication via email. " Manually, complete and return the enclosed Response Form. If you have any questions regarding this notice, please contact our Customer Support: " From our website: http://www.beckmancoulter.com
Quantity
1 system
Distribution
Worldwide distribution - US Nationwide.
Lot, serial or model codes
UDI-DI: 15099590369231; Serial Numbers: 608109
510(k) numbers
["K023764"]
PMA numbers
not on file
Product code
JJE
Firm FEI
2122870
Firm city
Chaska
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register