FDA Medical Device Recalls (openFDA)
Intelerad Medical Systems: IntelePACS software versions 3-8-1 to 4-9-1, Product Usage: IntelePACS is a device that receives digital images and data from various sources (such as, CT scanners, MRI scanners, ultrasound systems, R/F units, computer and direct radiographic devices, secondary capture devices, scanners, imaging gateways, or other imaging sources.
- Recall number
- Z-2267-2014
- Recalling firm
- Intelerad Medical Systems
- Product
- IntelePACS software versions 3-8-1 to 4-9-1, Product Usage: IntelePACS is a device that receives digital images and data from various sources (such as, CT scanners, MRI scanners, ultrasound systems, R/F units, computer and direct radiographic devices, secondary capture devices, scanners, imaging gateways, or other imaging sources.
- Status
- Terminated
- Initiated
- 2014-07-03
- Posted
- 2014-08-19
- Terminated
- 2016-01-13
- Reason
- In specific situation a partially truncated file may be written to the IntelePACS server. This partially truncated file still can be viewed by InteleViewer and InteleViewer will report to the user that the file in question is complete. This situation only occurs if a file is partially truncated.
- Root cause
- Software design
- Action
- Intelerad sent an Urgent Recall for Product Correction letter to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were informed that a new software release will be installed on the system. Afterwards, an audit will be run on the the entire PACS to identify all image data that may have been truncated. Customers were requested to complete and return the attached Client Response Form.
- Quantity
- 230
- Distribution
- Worldwide Distribution - USA Nationwide and the countries of Scotland, Ireland, United Kingdom, Canada, Australia, and New Zealand.
- Lot, serial or model codes
- software versions 3-8-1 to 4-9-1
- 510(k) numbers
- ["K083520"]
- PMA numbers
- not on file
- Product code
- LLZ
- Firm FEI
- 3008437943
- Firm city
- Montreal
- Firm state
- not on file
- Firm country
- Canada
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28