Skip to the content

FDA Medical Device Recalls (openFDA)

Vascutek, Ltd.: Gelweave (TM) Vascular Prostheses Manufactured by - VASCUTEK Ltd a TERUMO company

Recall number
Z-2266-2014
Recalling firm
Vascutek, Ltd.
Product
Gelweave (TM) Vascular Prostheses Manufactured by - VASCUTEK Ltd a TERUMO company
Status
Terminated
Initiated
2014-07-15
Posted
2014-08-18
Terminated
2015-03-19
Reason
The cautery that is provided with the Gelweave graft in the box has a shorter shelf life indicated on its label, versus that of the graft (4 years and 5 years, respectively).
Root cause
Incorrect or no expiration date
Action
Vascutek sent an Urgent Field Safety Notice dated July 15, 2014, to all affect customers. The letter identified the product the problem and the action needed to be taken by the customer. User Instructions: A. Vascutek advises that you ensure suitable equipment is available prior to implantation to trim the graft if required. B. For product received with a cautery in the interim, Vascutek request that you check the shelf life of each product before every operation. If you find that the cautery is out of date, remove and discard the cautery. If you require an appropriate cautery supplied as a separate item, please contact your local sales representative for a referral. C.Vascutek are taking this opportunity to reinforce the cautions from the Gelweave TM Vascular Prostheses IFU regarding soaking of the graft in saline to prevent focal burning if a cautery is to be used. This statement is already included in the IFU and now appears in bold, red font to ensure appropriate awareness. This notice needs to be passed on to all persons who need to be aware within your organization or to any organization where the devices have been transferred or distributed. Please maintain awareness of this Field Safety Notice while all actions are taken in your organization and until these products are provided without the cautery. Please return the User Return Slip by e-mail or fax to the Distributor's address given on page 3. This action by Vascutek Ltd. is being taken with the knowledge of the National Competent Authority Medicines and Healthcare Products Regulatory Agency (MHRA). If you have any further questions or comments, please do not hesitate to contact us at tcvs.recall@terumomedical.com. Return Completed Form immediately fax to (734) 741-6149. If required, your Vascutek Ltd. representative can discuss and provide more information on the use of cautery with Vascutek products. For further questions call (800) 262-3304 ext. 6056
Quantity
63,162
Distribution
Nationwide Distribution including the states of : AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI and WV.
Lot, serial or model codes
731030, 731032, 731034, 731036,731038, 731206, 731218, 731220, 731222, 731224, 731226, 731228, 731407, 731506, 731507, 731508, 731509, 731510, 731512, 731514, 731516, 731608, 731809, 732010, 732030, 732032, 732034, 732038, 732211, 732412, 732518, 732520, 732522, 732524, 732526, 732528, 733006, 733007, 733008, 733010, 733012, 733014, 733016, 733018, 733020, 733022, 733024, 733026, 733028, 733030, 733032, 733034, 733036, 733038, 733040, 733042, 734030, 734032, 734034, 734036, 734038, 735018, 735020, 735022, 735024, 735026, 735028, 736006, 736007, 736008, 736010, 736012, 736014, 736016, 736018, 736020, 736022, 736024, 736026, 736028, 736030, 736032, 736034, 736036, 736038, 731206/10/10, 731206/10/10RM, 731206/7, 731407/10/10, 731407/10/10RM, 731407/6X2RM, 731407/7, 731608/10/10, 731608/10/10RM, 731608/6, 731608/7, 731608/7X2RM, 731608/8/10, 731608/8/10RM, 731809/6, 732010/6, 732412/6, 7330303BL, 7330323BL, 7330343BL, 7330363BL, 7330383BL
510(k) numbers
not on file
PMA numbers
not on file
Product code
DSY
Firm FEI
3002808437
Firm city
Inchinnan
Firm state
not on file
Firm country
United Kingdom

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register