Skip to the content

FDA Medical Device Recalls (openFDA)

TriMed Inc.: Nitinol Staple, Elastic Staple Small-Symmetrical, Sterile R, Model #: a) ES-08x08; b) ES-10x10; c) ES-10x13; d) ES-12x10; e) ES-15x12; f) ES-10x14; g) ES-10x15

Recall number
Z-2265-2018
Recalling firm
TriMed Inc.
Product
Nitinol Staple, Elastic Staple Small-Symmetrical, Sterile R, Model #: a) ES-08x08; b) ES-10x10; c) ES-10x13; d) ES-12x10; e) ES-15x12; f) ES-10x14; g) ES-10x15
Status
Terminated
Initiated
2018-04-13
Posted
not on file
Terminated
2021-01-22
Reason
During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with the regulations could not be found. Sterilization records indicate that all lots of staples were irradiated to at least the defined dose. There is a risk that this issue could result in patient infection if not properly sterilized.
Root cause
Process design
Action
On 04/13/18, customer recall communication forms were sent via mail informing customers to quarantine effected devices, call the manufacturer to arrange device return at 800-633-7221, and report adverse events to the FDA's MedWatch.
Quantity
1,194
Distribution
U.S.: IL, UT, PA, NJ, WA, TX, NV, CA, NC, MS, KS, MO, MN, FL; OUS: Australia and Sweden
Lot, serial or model codes
Lot #: a) 021C, 00T3, 050501, 10505-01, 00501, 179201, 17920, 179202, 179203; b) 26401, 0219, 02613, 0216, 026B, 050502, 10505-02, 01P2, 24601, 179302, 179301, 179303, 264-01, 179304, 317003; c) R884, 0266, 050503, 10505-03, 179501, 004J, 00A0, 173501; d) 26402, 264-02, 2015, 050508, 10505-08, 24602, 11799, 23402, 0215, 1799, 179901, 179304, 1179; e) 26403, 03UQ, 050509, 10505-09, 24603, 11802, 1180202, 180202, 80202, 180203, 108202, 263401, 23401; f) 004K, 0093; g) 004L, 2012BB
510(k) numbers
["K112794"]
PMA numbers
not on file
Product code
JDR
Firm FEI
3001236812
Firm city
Santa Clarita
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register