Endologix, Inc.: Ovation iX Abdominal Stent Graft Systems, REF: TV-AB2080-I, TV-AB2080-J, TV-AB2380-I, TV-AB2380-J, TV-AB2680-I, TV-AB2680-J, TV-AB2980-I, TV-AB2980-J, TV-AB3480-I, TV-AB3480-J

Recall number
Z-2263-2020
Recalling firm
Endologix, Inc.
Product
Ovation iX Abdominal Stent Graft Systems, REF: TV-AB2080-I, TV-AB2080-J, TV-AB2380-I, TV-AB2380-J, TV-AB2680-I, TV-AB2680-J, TV-AB2980-I, TV-AB2980-J, TV-AB3480-I, TV-AB3480-J
Status
Terminated
Initiated
2020-05-06
Posted
2020-06-11
Terminated
2024-05-03
Reason
A material weakness adjacent to the polymer fill channel may become compromised during pressurization with liquid polymer. Polymer leaks may occur during implantation of abdominal stent graft system. Clinical events related to polymer leaks may be systemic and/or aneurysm related.
Root cause
Process control
Action
On 05/06/20, "Urgent Medical Device Correction notices" were distributed to customers via person/mail/email. These notices reaffirmed treatment recommendations for patients provided in the previously distributed Urgent Field Safety Notice, dated 09/26/2018 that were mailed on 09/27/18. Safety Update: Treatment of a Patient with Polymer Leak - Patient Reaction During the polymer injection step of the procedure, systemic hypotension may indicate that a polymer leak is occurring. Blood pressure monitoring during polymer fill may assist in early identification of a potential polymer leak. In the absence of other obvious diagnoses causing sudden hypotension during polymer fill, firm recommends that a hypersensitivity reaction (an anaphylactoid response) to intravascular polymer leak be considered a probable diagnosis. Patients with a polymer leak should undergo immediate treatment for a potential severe hypersensitivity response in accordance with institutional protocols (e.g., intravascular fluids, antihistamines, corticosteroids, epinephrine). In addition to systemic hypotension, device related findings that are indicative of a polymer leak include rapid emptying of the fill polymer syringe, and incomplete filling of the polymer channels. Safety Update: Treatment of a Patient with Polymer Leak - Aneurysm Management Aneurysm related complications that may occur due to polymer leak should be treated with standard endovascular techniques at the physician's discretion, utilizing the ancillary equipment listed in the Instructions for Use (IFU), or an open surgical approach. The specific treatment will be dependent on the extent and location of incomplete filling of the polymer rings and the associated clinical findings. In respect of intra-operative Type 1a endoleaks resulting from polymer leak, there were two main treatment strategies: (1) conservative management (in the cases of small endoleaks expected to resolve spontaneously) or (2) the use of balloon expand
Quantity
9156
Distribution
Worldwide DIstribution: U.S (nationwide).: PA, AZ, NY, FL, HI, CA, OR, IL, MS, AR, WI, KY, TX, TN, MO, NJ, MA, MT, VA, MI, CO, WV, NC, NH, NV, CT, ME, WY, OH, OK, IN, UT, AL, WA, GA, SC, KS, LA, NM, MD, DC, NE, SD, RI, IA, VT, ND. and Countries of:: Argentina, Australia, Chile, Hong Kong, Japan, Malaysia, Mexico, New Zealand, Taiwan, Thailand, Canada, Austria, Belgium, Cyprus, Czech Republic, Germany, Hungary, Ireland, Italy, Switzerland, Luxembourg, Netherlands, Poland, Portugal, Romania, Spain, Sweden, and United Kingdom
Lot, serial or model codes
All Lot and Serial Numbers not yet implanted
510(k) numbers
not on file
PMA numbers
["P120006"]
Product code
MIH
Firm FEI
3008011247
Firm city
Santa Rosa
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register