FDA Medical Device Recalls (openFDA)
Endologix, Inc.: Ovation iX Abdominal Stent Graft Systems, REF: TV-AB2080-I, TV-AB2080-J, TV-AB2380-I, TV-AB2380-J, TV-AB2680-I, TV-AB2680-J, TV-AB2980-I, TV-AB2980-J, TV-AB3480-I, TV-AB3480-J
- Recall number
- Z-2263-2020
- Recalling firm
- Endologix, Inc.
- Product
- Ovation iX Abdominal Stent Graft Systems, REF: TV-AB2080-I, TV-AB2080-J, TV-AB2380-I, TV-AB2380-J, TV-AB2680-I, TV-AB2680-J, TV-AB2980-I, TV-AB2980-J, TV-AB3480-I, TV-AB3480-J
- Status
- Terminated
- Initiated
- 2020-05-06
- Posted
- 2020-06-11
- Terminated
- 2024-05-03
- Reason
- A material weakness adjacent to the polymer fill channel may become compromised during pressurization with liquid polymer. Polymer leaks may occur during implantation of abdominal stent graft system. Clinical events related to polymer leaks may be systemic and/or aneurysm related.
- Root cause
- Process control
- Action
- On 05/06/20, "Urgent Medical Device Correction notices" were distributed to customers via person/mail/email. These notices reaffirmed treatment recommendations for patients provided in the previously distributed Urgent Field Safety Notice, dated 09/26/2018 that were mailed on 09/27/18. Safety Update: Treatment of a Patient with Polymer Leak - Patient Reaction During the polymer injection step of the procedure, systemic hypotension may indicate that a polymer leak is occurring. Blood pressure monitoring during polymer fill may assist in early identification of a potential polymer leak. In the absence of other obvious diagnoses causing sudden hypotension during polymer fill, firm recommends that a hypersensitivity reaction (an anaphylactoid response) to intravascular polymer leak be considered a probable diagnosis. Patients with a polymer leak should undergo immediate treatment for a potential severe hypersensitivity response in accordance with institutional protocols (e.g., intravascular fluids, antihistamines, corticosteroids, epinephrine). In addition to systemic hypotension, device related findings that are indicative of a polymer leak include rapid emptying of the fill polymer syringe, and incomplete filling of the polymer channels. Safety Update: Treatment of a Patient with Polymer Leak - Aneurysm Management Aneurysm related complications that may occur due to polymer leak should be treated with standard endovascular techniques at the physician's discretion, utilizing the ancillary equipment listed in the Instructions for Use (IFU), or an open surgical approach. The specific treatment will be dependent on the extent and location of incomplete filling of the polymer rings and the associated clinical findings. In respect of intra-operative Type 1a endoleaks resulting from polymer leak, there were two main treatment strategies: (1) conservative management (in the cases of small endoleaks expected to resolve spontaneously) or (2) the use of balloon expand
- Quantity
- 9156
- Distribution
- Worldwide DIstribution: U.S (nationwide).: PA, AZ, NY, FL, HI, CA, OR, IL, MS, AR, WI, KY, TX, TN, MO, NJ, MA, MT, VA, MI, CO, WV, NC, NH, NV, CT, ME, WY, OH, OK, IN, UT, AL, WA, GA, SC, KS, LA, NM, MD, DC, NE, SD, RI, IA, VT, ND. and Countries of:: Argentina, Australia, Chile, Hong Kong, Japan, Malaysia, Mexico, New Zealand, Taiwan, Thailand, Canada, Austria, Belgium, Cyprus, Czech Republic, Germany, Hungary, Ireland, Italy, Switzerland, Luxembourg, Netherlands, Poland, Portugal, Romania, Spain, Sweden, and United Kingdom
- Lot, serial or model codes
- All Lot and Serial Numbers not yet implanted
- 510(k) numbers
- not on file
- PMA numbers
- ["P120006"]
- Product code
- MIH
- Firm FEI
- 3008011247
- Firm city
- Santa Rosa
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28