FDA Medical Device Recalls (openFDA)

Ventana Medical Systems Inc: Model Number: 06917259001 UDI: 04015630976010 - Product Usage: The System s intended use environment includes anatomic pathology (AP) laboratories and histology facilities within hospital, clinic, reference, or private lab settings. The System is to be operated by qualified users in

Recall number
Z-2262-2020
Recalling firm
Ventana Medical Systems Inc
Product
Model Number: 06917259001 UDI: 04015630976010 - Product Usage: The System s intended use environment includes anatomic pathology (AP) laboratories and histology facilities within hospital, clinic, reference, or private lab settings. The System is to be operated by qualified users in
Status
Completed
Initiated
2019-10-04
Posted
not on file
Terminated
not on file
Reason
The firm became aware of a cleaning solution leaking issue from the middle staining module of the Ventana HE 600 instrument, used for histologic section samples in anatomic pathology laboratory settings, which caused fluid to travel down to the electrical wiring connector of the Coverslipper solvent shield circuit board. This can cause an electrical short and result in burned components.
Root cause
Under Investigation by firm
Action
All US consignee accounts who are utilizing the VENTANA HE 600 system were to receive a copy of the Urgent Medical Device Correction (UMDC) notification, VENTANA HE 600 Instrument - Fluid Leak and Risk of Burned Components, via UPS Ground on 10/16/19. It notified user that Roche Tissue Diagnostics received notification of an issue on the VENTANA HE 600 instrument. A leak from the middle staining module caused fluid to travel down to the electrical wiring connector of the Coverslipper solvent shield circuit board. This caused an electrical short and resulted in burned components. It stated clinical significance that there is a remote possibility that use of the VENTANA HE 600 system would result in adverse health consequences. It advised users that Roche Service Representatives will remove the cable that conducts electricity to the Coverslipper Module LED PCBA, proactively replace the Stainer Exchange Valves, and proactively replace the 65 micron filters in the Cleaning Solution fluidics line. Site visits will be scheduled as appropriate. Customers can continue using the instrument as normal. However, if droplets are observed on the instrument Coverslip Module during normal operation: stop instrument operation, unplug the instrument, and contact your Roche Service Representative immediately. Actions Required: " If customers observe droplets on the instrument Coverslip Module during normal operation: stop instrument operation, unplug the instrument, and contact your Roche Service Representative immediately. " Complete attached fax form and fax or email it according to the instructions on the form. " File the Urgent Medical Device Correction (UMDC) for future reference. Customers can contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-227-2155 if they have questions about the information contained in the UMDC. Adverse events or quality problems experienced with the use of this product may also be reported to
Quantity
461
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, MD, MI, MA, MN, MO, MS, MT, NC, NE, NJ, NY, OH, OK, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI and the countries of Australia, Austria, Belgium, Brazil, Canada, Chile, China, Columbia, Costa Rica, Croatia, Denmark, Ecuador, El Salvador, Finland, France, Germany, Hong Kong, Ireland, Italy, Japan, Kuwait, Lebanon, Malaysia, Norway, Peru, Poland, Saudi Arabia, Slovakia, Slovenia, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom, Vietnam.
Lot, serial or model codes
All lots
510(k) numbers
not on file
PMA numbers
not on file
Product code
KPA
Firm FEI
1000125183
Firm city
Oro Valley
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register