Medline Industries, LP: Medline medical convenience kit labeled as EYE TRAY-LF, Medline Kit SKU DYNJ21627P
- Recall number
- Z-2260-2026
- Recalling firm
- Medline Industries, LP
- Product
- Medline medical convenience kit labeled as EYE TRAY-LF, Medline Kit SKU DYNJ21627P
- Status
- Open, Classified
- Initiated
- 2026-04-10
- Posted
- 2026-05-28
- Terminated
- not on file
- Reason
- Specific Medline Kits contain Cardinal Health Monoject 1mL Luer Lock Syringes, Tuberculin Print, Sterile. Cardinal Health has recalled these syringes because the outer carton and blister pack are labeled as a 1 mL Luer Lock Tuberculin Syringe (Product Code 1180100777), but the syringes inside are U-100 insulin syringes.
- Root cause
- Under Investigation by firm
- Action
- Medline Industries, LP issued an URGENT MEDICAL DEVICE RECALL notice to it consignees on 4/10/2026 via USPS first class mail and email. The notice explained the issue, potential risk, and requested the following: "REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed on the enclosed response form. Quarantine all affected product. 2. Please complete and return the enclosed response form listing the quantity of affected product you have in inventory. Even if you do not have any affected product in inventory, please complete and return the response form. 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product." For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
- Quantity
- 8 kits
- Distribution
- US Nationwide distribution in the states of California.
- Lot, serial or model codes
- UDI/DI 10198459151569 (each), 40198459151560 (case), Lot Numbers: 26BMC538
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- OJK
- Firm FEI
- 1417592
- Firm city
- Northfield
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28