Skip to the content

FDA Medical Device Recalls (openFDA)

GE Healthcare, LLC: System, Tomography Computed Emmission

Recall number
Z-2259-2020
Recalling firm
GE Healthcare, LLC
Product
System, Tomography Computed Emmission
Status
Open, Classified
Initiated
2018-11-30
Posted
2020-06-03
Terminated
not on file
Reason
GE Healthcare has identified that a small number of Nuclear Medicine systems are operating without currently updated software. In a recent event, a patient s finger was injured during the Unload process while the automatic Unload feature was being used; the software on the system was not current.
Root cause
Other
Action
GE Healthcare identified that a small number of Nuclear Medicine systems are operating without currently updated software.
Quantity
526 total units
Distribution
Domestic Distribution: AL, AR, AZ, CA, CO, CT, DC, DE, FL,GA, IA, IL, IN, KY, LA,MA, MD, ME, MI, MN, MO, MS, NC, NE, NM, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT,VA, WI and WV. Foreign Distribution: Argentina, Australia, Belgium, Brazil, Canada, Colombia, Denmark, France, Germany, Ghana, India, Israel, Italy, Japan, Korea, Malaisya, Mexico, Netherlands, New Zealand, Norway, Poland, Portugal, Singapore, Sweden, Taiwan, Thailand, Turkey and United Kingdom.
Lot, serial or model codes
Model for Discovery NM 530c: 5248934, 5166674-01, H3000SA, H3001JA, H3001JD.  Serial Numbers:  19016,0000000ME11041,19069,19098,19082,19070,19086,19027,19044,19083,19065,19062,19048,19009,19109,19015,19040,19085,19012, 19108, 19030, 19057,000000109961-1,19002,16001,16007,19008,19097,19072,19080,16020,19026,19005,19025,0000000000ME-4.
510(k) numbers
["K080124"]
PMA numbers
not on file
Product code
KPS
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register