FDA Medical Device Recalls (openFDA)
Quantimetrix Corporation: Dropper Plus Point-of-Care Urinalysis Dipstick Control kit The Quantimetrix Dropper Plus Point-of-Care Urinalysis Dipstick Control is intended as a control for urinalysis reagent strips, microalbumin, and creatinine by the listed test methods, and as a control for confirmatory tests such as K-CHECK and Ictotest reagent tablets, and for hCG methods.
- Recall number
- Z-2257-2018
- Recalling firm
- Quantimetrix Corporation
- Product
- Dropper Plus Point-of-Care Urinalysis Dipstick Control kit The Quantimetrix Dropper Plus Point-of-Care Urinalysis Dipstick Control is intended as a control for urinalysis reagent strips, microalbumin, and creatinine by the listed test methods, and as a control for confirmatory tests such as K-CHECK and Ictotest reagent tablets, and for hCG methods.
- Status
- Terminated
- Initiated
- 2017-12-21
- Posted
- not on file
- Terminated
- 2021-01-27
- Reason
- The kits contained an incorrect lot of Dropper Plus Level 1 vials
- Root cause
- Mixed-up of materials/components
- Action
- Communication with their only customer for this recall was via e-mail and phone. Upon notification by the customer the following activities were initiated: All lots of the Dropper Plus product in inventory were 100% inspected. The customer was ask to return all of the kits to Quantimetrix. For further questions please call (310) 536-0006.
- Quantity
- 20 kits
- Distribution
- US Distribution to the state of : Kansas
- Lot, serial or model codes
- Lot 44450E
- 510(k) numbers
- ["K874890"]
- PMA numbers
- not on file
- Product code
- JJW
- Firm FEI
- 2020715
- Firm city
- Redondo Beach
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28