Skip to the content

FDA Medical Device Recalls (openFDA)

Arrow International Inc: Arrow VPS Access Kit with Maximal Barrier Precautions for use with 5 Fr. Peripherally inserted Central Venous Catheter The Arrow(r) CVC is indicated to provide short-term (<30 days) central venous access for treatment of diseases or conditions requiring central venous access including, but not limited to: multiple infusions of fluids, medications, or chemotherapy; infusion of fluids that are hypertonic, hyperosmolar, or have divergent pH values; frequent blood sampling or blood/blood component infusions; infusion of incompatible medications; central venous pressure monitoring; lack of usable peripheral IV sites; replacement of multiple peripheral sites for IV access.

Recall number
Z-2257-2017
Recalling firm
Arrow International Inc
Product
Arrow VPS Access Kit with Maximal Barrier Precautions for use with 5 Fr. Peripherally inserted Central Venous Catheter The Arrow(r) CVC is indicated to provide short-term (<30 days) central venous access for treatment of diseases or conditions requiring central venous access including, but not limited to: multiple infusions of fluids, medications, or chemotherapy; infusion of fluids that are hypertonic, hyperosmolar, or have divergent pH values; frequent blood sampling or blood/blood component infusions; infusion of incompatible medications; central venous pressure monitoring; lack of usable peripheral IV sites; replacement of multiple peripheral sites for IV access.
Status
Terminated
Initiated
2017-03-20
Posted
not on file
Terminated
2018-12-18
Reason
Arrow International initiated the recall due to reports of safety cover disengagement and needlestick injury for the BD Eclipse Needle that are packaged with certain Arrow products.
Root cause
Under Investigation by firm
Action
Arrow International sent an Urgent Medical Device Notification Letter dated April 5, 2017, to affected customers to inform them of the issue. The notice provided safety instructions for using the product and instructed customers to return the enclosed acknowledgement form indicating the notice was received, and fax it to 855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. Customers with questions were instructed to call 1-866-246-6990. For questions regarding this recall call 610-378-0131.
Quantity
43,394 units
Distribution
Nationwide Distribution
Lot, serial or model codes
Material Number ASK-04001-DU7
510(k) numbers
["K002507"]
PMA numbers
not on file
Product code
FOZ
Firm FEI
3015859709
Firm city
Reading
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register