Siemens Medical Solutions USA, Inc.: The following Pro.specta Emission Computed Tomography Systems that were manufactured with an SLD Look Ahead Optical Sensor Assembly: (a) Symbia Pro.specta Q3, Model 11364751; (b) Symbia Pro.specta X3, Model 11364752; (c) Symbia Pro.specta X7, Model 11364753.

Recall number
Z-2256-2023
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
The following Pro.specta Emission Computed Tomography Systems that were manufactured with an SLD Look Ahead Optical Sensor Assembly: (a) Symbia Pro.specta Q3, Model 11364751; (b) Symbia Pro.specta X3, Model 11364752; (c) Symbia Pro.specta X7, Model 11364753.
Status
Open, Classified
Initiated
2023-07-12
Posted
2023-07-28
Terminated
not on file
Reason
During a whole-body planar scan with auto-contour, there is a possibility the short-linear drive (SLD) look-ahead sensors may trigger earlier than expected and may result in detector 1 positioning further from the patient than needed leading to a reduction in image resolution.
Root cause
Device Design
Action
The firm issued undated letters flagged Customer Safety Advisory Notice CAN 001-2023 on 7/12/2023 via email, certified mail, or, if necessary, hand delivery. The letter explains the issue, when the malfunction occurs, the potential risks, and how the consignee can help to avoid the potential risk of the issue. The letter informs the consignee that Siemens Healthineers is working to address the issue through a scheduled service visit and the consignee's local service organization will begin contacting the consignees in the 4th quarter of 2023 to schedule this work. In the meantime, the notice is to be placed in the Symbia Pro.specta Operator Manual and disseminated to all operators of the scanner. No response form was enclosed. The recalling firm uses an SAP system for tracking purposes. For the letters that are delivered via email, the consignee verifies receipt through an adobe signature. For the letters delivered via certified mail, there is a return receipt.
Quantity
108 systems
Distribution
Distribution was made to ID, KS, MA, NH, OH, TX, and VA. There was no government/military distribution. Foreign distribution was made to Argentina, Australia, Austria, Belgium, Canada, Colombia, Denmark, Finland, France, Germany, Hong Kong, Israel, Italy, Japan, Netherlands, Portugal, Singapore, Switzerland, Turkey, United Arab Emirates, and United Kingdom.
Lot, serial or model codes
Software version VA20A. (a) Serial numbers 100201, 100202, 100203, 100204, 100205, 100206, 100207, 100209, 100208, 100210, 100212, 100211, 100213, 100215, 100217, 100219, 100218, 100220, 100221, and 100224; UDI-DI 4056869264745. Additional serial numbers as of 09/07/2023: 100223, 100225, 100227, 100228, 100237; Additional serial numbers as of 10/17/2023: 100214, 100229; Additional serial numbers as of 12/12/2023: 100232, 100231, 100234; Additional serial number as of 01/15/2024:100226. (b) Serial numbers 100101, 100102, 100201, 100203, 100204, 100207, 100206, 100208, 100205, 100209, 100210, 100212, 100213, 100214, 100215, 100216, 100218, 100217, 100219, 100220, 100221, 100223, 100224, 100222, 100226, 100228, 100227, 100225, 100229, 100231, 100232, 100233, 100235, 100234, 100237, 100239, 100241, 100240, 100236, 100238, 100242, 100243, 100247, 100244, 100250, 100251, 100252, and 100255; UDI-DI 4056869264752; Additional serial numbers as of 09/07/2023: 100245, 100256, 100257, 100258, 100261, 100262, 100265, 100273; Additional serial numbers as of 10/17/2023: 100230, 100266; Additional serial numbers as of 12/12/2023: 100269; Additional serial numbers as of 01/15/2024: 100248, 100254, 100260, 100271, 100272. (c) Serial numbers 100101, 100201, 100203, 100202, 100205, and 100206; UDI-DI 4056869264769; Additional serial numbers as of 09/07/2023: 100208, 100210; Additional serial numbers as of 12/12/2023: 100209
510(k) numbers
["K212604"]
PMA numbers
not on file
Product code
KPS
Firm FEI
1000514697
Firm city
Hoffman Estates
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register