Skip to the content

FDA Medical Device Recalls (openFDA)

Ortho Development Corporation: Tibila Insert UC, Balanced Knee System, REF163-3YYY Product Usage: UHMWPE component used in TKA procedures to serve as a tibial bearing surface for articulation with the femoral component

Recall number
Z-2256-2014
Recalling firm
Ortho Development Corporation
Product
Tibila Insert UC, Balanced Knee System, REF163-3YYY Product Usage: UHMWPE component used in TKA procedures to serve as a tibial bearing surface for articulation with the femoral component
Status
Terminated
Initiated
2014-07-11
Posted
2014-08-17
Terminated
2014-10-06
Reason
Ortho Development is recalling various Pivot Bipolar Cup, Patella, PS Tibial Insert, UC Tibila Insert, and CK Tibil Insert because the foil pouches did not properly seal, allowing air to seep back into the pouch after sealing which increases the risk that the material properties of the implant would be substantially degraded due to oxidation.
Root cause
Packaging process control
Action
Ortho Development notified all affected consignees/distributors of the recall via email on July 11, 2014 which included a list of the affected product (part numbers, lot numbers and quantities) under their control. An RMA number and return shipping label were also included with the email. Emails were followed up with phone calls to verify receipt of the notification.
Quantity
13
Distribution
Worldwide Distribution - US Nationwide and the country of Japan. No military/govt/VA consignees.
Lot, serial or model codes
UC Tibial Insert Sz 3 10mm,Part 163-3310,Lot A136109,UC Tibial Insert Sz 4 8mm,Part 163-3408,Lot A135948,UC Tibial Insert Sz 4 14mm,Part 163-3414,Lot A136110,UC Tibial Insert Sz 6 7mm,Part 163-3607,Lot A136113,UC Tibial Insert Sz 6 7mm,Part 163-3607,Lot A136746,
510(k) numbers
["K090705"]
PMA numbers
not on file
Product code
JWH
Firm FEI
3017298928
Firm city
Draper
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register