Cyberonics, Inc: VNS Therapy Programmer, Model 3000, v1.0 System

Recall number
Z-2255-2018
Recalling firm
Cyberonics, Inc
Product
VNS Therapy Programmer, Model 3000, v1.0 System
Status
Terminated
Initiated
2017-11-13
Posted
not on file
Terminated
2022-10-31
Reason
Certain Model 3000 programming events can result in miscalculation of parameters stored in the Models 103, 104, 105, and 106 generators. During these programming events, the miscalculations can lead to: " Delivery of more stimulation than intended, resulting in painful stimulation or other common side effects (Model 106 only); " No stimulation in the case of device disablement (Burst Watchdog Timeout), resulting in no therapy to the patient (Model 106 only); " Delivery of less stimulation than intended, resulting in therapeutic settings not being achieved within device specification (Models 103, 104, 105, or 106); and/or " Delays or absence of the 75% and 50% battery life indicators displayed by the programming software (Models 103, 104, 105, or 106).
Root cause
Device Design
Action
On November 13, 2017 LivaNva sent letters to all the doctor and surgeons that use this device. "Dear Doctor: This notification describes an issue identified with the VNS Therapy Programmer, Model 3000 version 1.0 which may impact patients implanted with Models 103, 104, 105 or 106 VNS Therapy generators. This issue does not affect Model 1000 SenTiva or Model 102/102R generators. What is the Issue? Certain programming events performed with a Model 3000 Programmer on a Model 103, 104, 105 or 106 VNS Therapy generator can result in a miscalculation of some parameters stored in the generator, which could lead to: " Output current delivered being different than that programmed; " Device disablement; and/or " Errors in the 75% and 50% battery life indicators. The actual battery life and battery function are not affected by this issue. Risk to Health This issue presents the following risks: " Delivery of more stimulation than intended, resulting in painful stimulation or other common side effects 1 (Model 106 only); " No stimulation in the case of device disablement, resulting in no therapy to the patient (Model 106 only); " Delivery of less stimulation than intended, resulting in therapeutic settings not being achieved (Models 103, 104, 105 or 106) ; or " Sudden decrease from 100% to 25% battery life indicator, though the actual battery life and battery function are not affected (Models 103, 104, 105). Which Patients are Potentially Impacted? Any patient with a Model 103, 104, 105 or 106 generator that has been programmed using a Model 3000 programmer may be affected by this issue. Your LivaNova representative will collect data from your Model 3000 programmer to identify which patients are impacted by this notice. Actions to be taken by the Physician 1. Stop using the Model 3000 programmer to program Model 103, 104, 105 or 106 generators. You shall continue using the Model 3000 programmer with Model 1000 SenTiva generators and may continue to use it wi
Quantity
252
Distribution
AL, CA, CO, DE, FL, GA, ID, IL, IN, ME, MO, MS, NC, NJ, NY, PA, TN, TN, TX, UT, WA and WI
Lot, serial or model codes
Serial Numbers: FMT77F2, FKR57F2, FCR77F2, CJR77F2, BNT77F2, BNS77F2, 7XR77F2, 7XP57F2, 79R77F2, 90S77F2, 97Q57F2, 6NT77F2, 9GP57F2, 6KQ57F2,  B7Q57F2, 57P57F2, 55T77F2, J6R57F2, HWR77F2, 1G7G1F2, 1JR57F2,  9XP57F2, 4LR77F2, 4GT77F2, 4GR57F2, 2DT77F2, G2S77F2, FVQ77F2,  G5T77F2, FDQ57F2, F5R57F2, F2R57F2, F1Q57F2, GHS77F2, DVR77F2,  DNP57F2, DMYB7F2, DJS77F2, D1R57F2, CVQ57F2, H4R57F2, 85S77F2,  89T77F2, 8LYB7F2, 97YB7F2, 6MQ57F2, 6CQ57F2, 5HS77F2, JLQ57F2,  JGS77F2, 11R57F2, 1TQ57F2, 1YR77F2, 2DR77F2 FXP57F2 , FTR77F2, FPQ57F2, FNQ57F2, FLYB7F2, FFP57F2, GGT77F2, GSQ57F2, CMS77F2, H1R77F2, H3R57F2, 7LS77F2, 9KR57F2, 9NS77F2, 65R57F2, B8R77F2,  B9T77F2, 5D0C7F2, B9YB7F2, BJR57F2, JCQ57F2, J9YB7F2, HMYB7F2, CDQ57F2, FXQ57F2, G5S77F2, G3R77F2, FRQ57F2, G7YB7F2, FLS77F2, G90C7F2, FGR77F2, FFT77F2, G9Q57F2, FCT77F2, G9R57F2, F9YB7F2,  F9S77F2, F8T77F2, GDQ57F2, F7R57F2, F4T77F2, DWR77F2, DTR77F2,  DSQ57F2, GMQ57F2, DLS77F2, DH0C7F2, DGS77F2, DGP57F2, GZR77F2, D1Q57F2, CSQ57F2, CLS77F2, C9YB7F2, C70C7F2, H1R57F2, C1R77F2,  7MS77F2, 78YB7F2, 9SQ57F2, 5VQ57F2, 5RQ57F2, B60C7F2, 5NQ57F2, 5MR77F2, 5KS77F2, B8T77F2, BHP57F2, BJR77F2, 4JS77F2, H4T77F2,  JZQ57F2, JMS77F2, JMQ57F2, JG0C7F2, JBQ57F2, J8T77F2 HZQ57F2, J7R57F2, HVR77F2, HTP57F2, HNS77F2, HMQ57F2, HKQ57F2, H7S77F2,  D3R57F2, 1CQ57F2, 1HT77F2, 1MYB7F2, 2CP57F2, 2LYB7F2, CLYB7F2,  CMYB7F2, B4R77F2, B2S77F2, B0R57F2, 9QS77F2, D8Q57F2, 960C7F2, 95T77F2, 8JR77F2, 8HR57F2, BXP57F2, BWR77F2, BTQ77F2, C8YB7F2,  C60C7F2, BLR77F2, BLQ57F2, BGR77F2, BGP57F2, CHQ57F2, CKR57F2, CPP57F2, CSR77F2, CTQ77F2, CXR77F2, 9MS77F2, 9JR77F2, 9DQ57F2, D7S77F2, 94T77F2, 8PQ57F2, BWQ77F2, BSR77F2, BNQ57F2, C9R57F2, C8R57F2, C5R57F2, CF0C7F2, C3R57F2, CFP57F2, BDT77F2, B9S77F2, B5R57F2, CPQ57F2, B4R57F2, D2R57F2, 9KQ57F2, D5R77F2, 9F0C7F2, 980C7F2, D9YB7F2, 8WR77F2, DB0C7F2, 9P7G1F2, D3S77F2, D6R57F2, D7R57F2, 9CR77F2, 9G0C7F2, 96R57F2, DFP57F2, C0S77F2, BN7G1F2, BQ7G1F2, C8S77F2, CB0C7F2, CCR77F2, BHR57F2, CHP57F2, BCR77F2, BBR77F2, BB0C7F2, B90C7F2, B70C7F2, B2R57F2, 9VP57F2, 9TQ57F2, 9GR57F2, 9FP57F2, 9BS77F2 98R57F2 ,D80C7F2, 970C7F2, D8YB7F2, 95R77F2, 8YR77F2, C1R57F2, C80C7F2, C3R77F2, CNS77F2, 9VQ77F2, D4T77F2, 9DR77F2, 91R57F2, 8VQ77F2, 8RQ57F2, DBS77F2, 8JS77F2, C0R57F2, BTR77F2, BMYB7F2, BKR77F2, BBP57F2 and B8S77F2
510(k) numbers
not on file
PMA numbers
["P970003"]
Product code
MUZ
Firm FEI
1000117004
Firm city
Houston
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register