Skip to the content

FDA Medical Device Recalls (openFDA)

Olympus Scientific Solutions Americas: Olympus Scientific Solutions Americas Corporation (OSSA) Model: Analytical X-Ray System X -Ray Fluorescence. Intended to emit ionizing radiation Delta with models: DC-2000, DC-4000, DC-6000, DI-2000, DP-2000, DP-4000, DP-4050, DP- 6000, DPO-4000, DPO-6000, DS-2000, DS-4000, DS-6000, DS-6500, Delta, Delta 50, Delta Classic Plus, Delta Element, Delta Premium , Delta Professional , and Delta-Pl us.

Recall number
Z-2255-2016
Recalling firm
Olympus Scientific Solutions Americas
Product
Olympus Scientific Solutions Americas Corporation (OSSA) Model: Analytical X-Ray System X -Ray Fluorescence. Intended to emit ionizing radiation Delta with models: DC-2000, DC-4000, DC-6000, DI-2000, DP-2000, DP-4000, DP-4050, DP- 6000, DPO-4000, DPO-6000, DS-2000, DS-4000, DS-6000, DS-6500, Delta, Delta 50, Delta Classic Plus, Delta Element, Delta Premium , Delta Professional , and Delta-Pl us.
Status
Open, Classified
Initiated
2016-06-22
Posted
not on file
Terminated
not on file
Reason
Trigger lock option does not always prevent x-ray generation. Specifically if the unit is in test mode or repeat test mode then the trigger lock may not prevent x-ray initiation when the trigger is pulled. The correction is via repair by a software update.
Root cause
Radiation Control for Health and Safety Act
Action
The firm notified customers via an "Important - Electronic Product Radiation Warning" letter. It described the issue; and, told customers to not rely on the Trigger Lock feature until the software upgrade had been completed. The letter also provided instructions for customers to not only determine their software version; but, also instructions to bring their affected device into compliance. The firm will remedy the issue without charge. Customers can contact their local representative or OSSA at ani.technicalsupport@olympus-ossa.com, phone: +1 866 446 6689, fax: +1 781 419 3980, with questions or if additional information is needed.
Quantity
2, 210 units distributed in the USA
Distribution
Nationwide Distribution.
Lot, serial or model codes
All Models of the Delta XRF Analyzer with Software version 2.5.17.1 and higher
510(k) numbers
not on file
PMA numbers
not on file
Product code
RBY
Firm FEI
3006241029
Firm city
Waltham
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register