FDA Medical Device Recalls (openFDA)
King Systems Corp.: King LT-D Oropharyngeal Airways The King LT-D is intended for use in adult patients (in excess of 25 kg) for controlled ventilation during anesthesia for procedures that are short in duration and when the patient is considered to have a low risk of aspiration of stomach contents.
- Recall number
- Z-2255-2013
- Recalling firm
- King Systems Corp.
- Product
- King LT-D Oropharyngeal Airways The King LT-D is intended for use in adult patients (in excess of 25 kg) for controlled ventilation during anesthesia for procedures that are short in duration and when the patient is considered to have a low risk of aspiration of stomach contents.
- Status
- Terminated
- Initiated
- 2013-09-03
- Posted
- 2013-09-19
- Terminated
- 2014-04-15
- Reason
- Report that size 5 KLTD airways were packaged in pouches indicating a size of 2.5 and shipped to your facility. The correct size of the KLTD airway is clearly marked on the actual airway product and the physical size difference between 2.5 and 5 is distinct and obvious.
- Root cause
- Labeling mix-ups
- Action
- King Systems notified one consignee who received the affected product verbally on August 7, 2013, via email and an Urgent Medical Device Recall - Immediate Action Required notification on September 3, 2013. The Urgent Medical Device Recall - Immediate Action Required notification described the product and issue as well as includes a response form for the consignee to complete and send back to King Systems via FAX to 317-776-5175, or email to wnet@ambu.com. Questions regarding the notice are directed to: King Systems customer service representative at 800.642.5464, Monday through Friday 8:00am- 5pm EDT. For questions regarding this recall call 317-776-6823.
- Quantity
- 9 cases; 45 products
- Distribution
- Distributed in NY.
- Lot, serial or model codes
- Part Number KLTD2125 Lot Number I10RG
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- CAE
- Firm FEI
- 1824226
- Firm city
- Noblesville
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28