Skip to the content

FDA Medical Device Recalls (openFDA)

Wilson-Cook Medical Inc.: Fusion Lithotripsy Extraction Basket, Model Number FS-LXB-2X4 G48277; Fusion Lithotripsy Extraction Baskets are intended to be used for endoscopic removal of biliary stones and foreign bodies.

Recall number
Z-2252-2023
Recalling firm
Wilson-Cook Medical Inc.
Product
Fusion Lithotripsy Extraction Basket, Model Number FS-LXB-2X4 G48277; Fusion Lithotripsy Extraction Baskets are intended to be used for endoscopic removal of biliary stones and foreign bodies.
Status
Open, Classified
Initiated
2023-06-22
Posted
2023-07-28
Terminated
not on file
Reason
Non-conforming product may appear in the field as inability to advance or retract the basket, or the drive wire breaking or separating from handle (with or without mechanical lithotripsy).
Root cause
Process control
Action
Cook Medical issued an URGENT: MEDICAL DEVICE REMOVAL notice to their consignees on 07/14/2023 by email or priority mail. The notice explained the issue, the risk, and requested the consignee cease use and return the affected product.
Quantity
393 units
Distribution
Worldwide distribution.
Lot, serial or model codes
UDI/DI 00827002482777, 10827002482774, Lot Numbers: W4589620, W4589141, W4596704, W4596704, W4589141, W4715238, W4596704, W4715238, W4715238, W4646708, W4644270, W4596704, W4589141, W4589141, W4715238, W4646708, W4715238, W4660182, W4596704, W4660182, W4660182, W4589141, W4596704, W4660182, W4715242, W4644270, W4589620, W4664573, W4646708, W4660182, W4589141, W4589620, W4644270, W4660182, W4660182, W4589620, W4644270, W4596704, W4644270, W4664573, W4644270, W4644270, W4646708, W4715238, W4664573, W4608745, W4670102, W4697072, W4697068, W4697072, W4696416, W4660182, W4620539, W4620539, W4620539, W4651764, W4651764, W4670097, W4664041, W4661895, W4647284, W4589621, W4693149, W4645756, W4594582, W4596701, W4612933, W4647284, W4664041, W4612933, W4643861, W4645763, W4594582, W4589621, W4635967, W4631138, W4644271, W4645308, W4645755, W4645756, W4647284, W4631138, W4596701, W4597312, W4612932, W4664041, W4594582, W4594582, W4597312, W4645755, W4661895, W4645763, W4571161, W4608304, W4610258, W4612932, W4612933, W4635967, W4645755, W4645755, W4693149, W4610258, W4596701, W4597312, W4645308, W4647284, W4635967, W4645308, W4693149, W4645763, W4571161, W4597312, W4612933, W4631138, W4631932, W4631932, W4571161, W4662353, W4635967, W4608304, W4612932, W4661895, W4645756, W4635967, W4635967, W4631932, W4635967, W4612933, W4631138, W4645308, W4596701, W4608304, W4662353, W4693149, W4589621, W4612933, W4645755, W4647284, W4662353, W4610258, W4643861, W4644271, W4644271, W4645308, W4645763, W4596701, W4608304, W4608304, W4612932, W4619650, W4622436, W4631932, W4635967, W4631138, W4594582, W4643861, W4631932, W4645756, W4662353, W4664041, W4594582, W4589621, W4664041, W4662353, W4612933, W4612933, W4631138, W4612933, W4631138, W4596704, W4589141
510(k) numbers
not on file
PMA numbers
not on file
Product code
FFL
Firm FEI
1037905
Firm city
Winston Salem
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register