Argon Medical Devices, Inc: Clear Flashback Needle 7cm x 18G AMC/4; Clear Flashback Needle 2.5 cm x 21G AMC/4; Winged Arterial Needle 7cm x 18G AMC/4; Clear Flashback Needle 7 cm x 19G AMC/4 Usage: Argon Arterial Access Needles are designed to provide vascular access for the percutaneous introduction of a guidewire.
- Recall number
- Z-2252-2017
- Recalling firm
- Argon Medical Devices, Inc
- Product
- Clear Flashback Needle 7cm x 18G AMC/4; Clear Flashback Needle 2.5 cm x 21G AMC/4; Winged Arterial Needle 7cm x 18G AMC/4; Clear Flashback Needle 7 cm x 19G AMC/4 Usage: Argon Arterial Access Needles are designed to provide vascular access for the percutaneous introduction of a guidewire.
- Status
- Terminated
- Initiated
- 2017-05-11
- Posted
- not on file
- Terminated
- 2018-01-25
- Reason
- Argon Medical has received a complaint from one of their distributors of a potential packaging defect. The product is sealed in a pouch and the defect if present shows as a potential void or unsealed edge of the pouch.
- Root cause
- Process design
- Action
- Argon Medical Devices, Inc. is sending an Urgent Medical Device Recall letter by registered mail to the affected customer/hospital/distributor. Customers are instructed to inspect and quarantine and return any affected product remaining at their facility utilizing the instruction on the Recall Response Form.
- Quantity
- 7140 pieces
- Distribution
- Worldwide Distribution - US Nationwide in the countries of Belgium, Canada, Japan, Netherlands
- Lot, serial or model codes
- 11161099 11161729 11160990 11160912
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- GDR
- Firm FEI
- 1625425
- Firm city
- Athens
- Firm state
- TX
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28