Skip to the content

FDA Medical Device Recalls (openFDA)

Circadiance LLC: SmartMonitor 2 PS/PSL, Monitor, Apnea, Facility USE. Intended for use in the continuous monitoring of respiration, heart rate, and SP02 of infant, pediatric and adult patients.

Recall number
Z-2250-2017
Recalling firm
Circadiance LLC
Product
SmartMonitor 2 PS/PSL, Monitor, Apnea, Facility USE. Intended for use in the continuous monitoring of respiration, heart rate, and SP02 of infant, pediatric and adult patients.
Status
Terminated
Initiated
2017-05-01
Posted
not on file
Terminated
2018-02-05
Reason
Circadiance has determined that it is possible for certain Smart Monitor 2PS/PSL monitors to exhibit intermittent operation of the nurse call feature. The firm has updated the device to increase the "Nurse Call" feature to ensure monitor alarms are transmitted to compatible nurse call systems.
Root cause
Under Investigation by firm
Action
Circadiance mailed a Medical Device Removal SmartMonitor 2 PS/PSL, dated 04/28/2017, to affected customers to inform them of the issue, the actions to be taken and the actions the firm plans to take to correct the issue. Customers are told to test the functionality of the system prior to each use; and, complete the Business Reply Form. The form can be returned via 412-202-4583 or emailed to james.gianoutsos@circadiance.com. Customers are to contact Circadiance Customer Service at 724-858-2837 to receive an Return Material Authorization number; or, 1-888-825-9640 to repair their unit free of charge, or for questions or support.
Quantity
1,242 units
Distribution
Nationwide Distribution.
Lot, serial or model codes
Unique Device Identifier (UDI) 1030271, 1038140, 1038141, U1030271, and U1038140.  All serial numbers of units with "REF", "MODEL" or standalone part number 1030271, 1038140, 1038141, U1030271 or U1038140 located on the products label are impacted by this notice. Monitors with the above part numbers that have a serial number starting with "CP" are NOT affected by this issue.
510(k) numbers
["K061256"]
PMA numbers
not on file
Product code
NPE
Firm FEI
3006182632
Firm city
Export
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register