FDA Medical Device Recalls (openFDA)
Becton Dickinson & Company: BDSpinal Tray with BDQuincke Needle 25 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Numbers: 405621, 405735.
- Recall number
- Z-2249-2026
- Recalling firm
- Becton Dickinson & Company
- Product
- BDSpinal Tray with BDQuincke Needle 25 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Numbers: 405621, 405735.
- Status
- Open, Classified
- Initiated
- 2026-04-27
- Posted
- 2026-05-29
- Terminated
- not on file
- Reason
- This is a sub-recall of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. The recall is due to quality issues identified during an FDA manufacturing facility inspection, citing specific quality deficiencies related to acceptance testing.
- Root cause
- Nonconforming Material/Component
- Action
- Becton Dickinson (BD) notified consignees via FedEx letter on about 04/27/2026. Customers were instructed to cease use of affected Bupivacaine ampules in the trays, secure an alternate local Bupivacaine drug prior to procedures, upon opening the kit, remove and destroy the affected Bupivacaine from the BD Anesthesia Tray, and if affected Bupivacaine was used with no issue there is no additional treatment or clinical follow up is needed, but if issues were encountered, report the incident per institutional guidelines and take any necessary actions. Additional instruction included to circulate the notification to applicable personnel or to any organizations where the potentially affected products have been transferred/distributed, post the notification on all inventory storage locations and any location where the product may be used, review recall letter BD received from Elevaris Medical Devices, and complete and return the provided Customer Response Form. Distributors were instructed to identify all affected inventory and attach the customer communication provided to each unit prior to distribution, identify customers within distribution network that purchased affected product, provide a copy of the customer letter to all customers to advise them of the field action notification on BD's behalf, circulate the notification to all applicable personnel, review recall letter BD received from Elevaris Medical Devices, and complete and return the provided Customer Response Form. BD has been made aware that the FDA has placed Bupivacaine manufactured by Huons Co., Ltd. on import alert #66-40 - Detention Without Physical Examination of Drugs From Firms Which Have Not Met Drug GMPs. Houns Co., Ltd. issued a product recall for its Bupivacaine ampules requesting quarantine of all ampules within expiry date. BD has ceased manufacturing kits/trays with the impacted Huons Co., Ltd Bupivacaine ampules. An updated notice was distributed to consignees on 7/8/26 clarifying recal
- Quantity
- 3,130 units
- Distribution
- Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, NC, NE, NH, NJ, NV, NY, OH, OR, PA, PR, SD, TN, TX, UT, VA, WA, WI, WV, WY.
- Lot, serial or model codes
- 1. Catalog Number: 405621. UDI Numbers: (01)00382904056216(17)260601(10)B01V142D, (01)00382904056216(17)260628(10)B01V176D. Lot Numbers (Expiration Date): B01V142D (2026-06-01), B01V176D (2026-06-28). 2. Catalog Number: 405735. UDI Numbers: (01)00382904057350(17)260601(10)B01V149D, (01)00382904057350(17)261001(10)B01V190D. Lot Numbers (Expiration Date): B01V149D (2026-06-01), B01V190D (2026-10-01).
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- CAZ
- Firm FEI
- 2243072
- Firm city
- Franklin Lakes
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28