Skip to the content

FDA Medical Device Recalls (openFDA)

Draeger Medical, Inc.: Carina Sub-Acute Care Ventilator

Recall number
Z-2248-2023
Recalling firm
Draeger Medical, Inc.
Product
Carina Sub-Acute Care Ventilator
Status
Open, Classified
Initiated
2023-07-12
Posted
2023-08-18
Terminated
not on file
Reason
Presence of polyether polyurethane (PE-PUR) and emission of 1,3-Dichloropropan-2-ol that exceed the acceptable uptake level during continuous use (>30 days) in pediatric patients.
Root cause
Nonconforming Material/Component
Action
The firm, Drager, sent an "Urgent Medical Device Recall" to customers by mail on 07/12/2023. The medical device recall notification instructs customers on how to utilize the Carina Ventilator until units can be corrected in the field by way of replacement of the blower to remove the presence of PE-PUR foam. Customers may continue to use the device until it has been modified provided that the set minute volume exceeds 3.6 l/min, only adult patients are ventilated, and only a Draeger ventilation hose with leak valve is used. The firm states that a new designed blower cover will be available in the early part of the fourth quarter of 2023 and Draeger will facilitate field service free of charge. Customers are to ensure all users and maintenance staff that utilize the recalled device are made aware of the recall notice. Customers with any questions regarding the operation of the Carina Ventilator are to contact Technical Support between 8:00 AM and 8:00 PM EST at 1-800-437-2437 (press 2, then 2, then 2); customers with any questions about the recall notice are to contact Michael Kelhart from 8:00 AM to 4:30 PM EST at 267-664-1131 or by email at mike.kelhart@draeger.com. The Medical Device Recall Response form is to be returned by fax 215-372-2940 or email: ditelford.quality@draeger.com to Draeger upon completion.
Quantity
11,621 units
Distribution
US Nationwide Distribution.
Lot, serial or model codes
Part No. 5704110; UDI-DI 04048675398516; All Serial No.
510(k) numbers
["K072885"]
PMA numbers
not on file
Product code
CBK
Firm FEI
2517967
Firm city
Telford
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register