Radiometer Medical ApS: safeCLINITUBES Blood Sampler;REF 942-898;D957P-70-70x1

Recall number
Z-2248-2020
Recalling firm
Radiometer Medical ApS
Product
safeCLINITUBES Blood Sampler;REF 942-898;D957P-70-70x1
Status
Terminated
Initiated
2020-03-17
Posted
not on file
Terminated
2022-07-28
Reason
The SafeCLINITUBES blood sampler may have insufficient aspiration capabilities ultimately resulting in lost samples.
Root cause
Process control
Action
The method of notification of the recall was a customer letter sent to all consignees via email and a hard copy sent via FedEx 2nd day with tracking and proof of delivery; date of issue was 3/17/20. Customers were instructed to do the following with the recalled product: -Cease using the Blood Samplers from the affected lots. -Check inventory of Blood Samplers from the affected lots. -Check for Blood Samplers from the affected lots distributed in the institution. -Collect any Blood Samplers from the affected lots. -Complete the Recall Response Form (last page of the customer advisory letter) and return it to the representative. -Discard the collected Blood Samplers of the affected lots. For any questions contact at 1-800-736-0600 option 1, Monday-Friday, 9am EST to 7pm EST.
Quantity
269 units
Distribution
U.S.: AK, AL, AR, CA, CO, CT, FL,GA, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OR, PA, SD, TN, TX, UT, VA, WA, and WI. OUS: Australia (New Zealand), Austria, Brazil Canada, China, Czech Republic (Slovakia), Denmark, Dubai (United Arab Emirates), Egypt, Finland, France, Germany, Hong Kong, Hungary, Iceland, India, Ireland, Italy, Japan, Kazakhstan, Lithuania, Malawi, Malaysia, Mauritius, Netherlands (Belgium), Nicaragua, Norway, Poland, Portugal, Russia, Saudi Arabia, Serbia, South Africa, Spain, Sweden, Switzerland, Turkey (Cyprus), and United Kingdom.
Lot, serial or model codes
Lot numbers: R1431 - R1495
510(k) numbers
not on file
PMA numbers
not on file
Product code
GIO
Firm FEI
3002807968
Firm city
Bronshoj
Firm state
not on file
Firm country
Denmark

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register