Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Communications: Berchtold Chromophare F300 Exam Light CHROMOPHARE examination lights are medical lights used for local illumination of the patients' body so that illnesses, injuries and disabilities can be diagnosed and treated

Recall number
Z-2248-2017
Recalling firm
Stryker Communications
Product
Berchtold Chromophare F300 Exam Light CHROMOPHARE examination lights are medical lights used for local illumination of the patients' body so that illnesses, injuries and disabilities can be diagnosed and treated
Status
Terminated
Initiated
2017-04-21
Posted
not on file
Terminated
2017-09-20
Reason
The exam light may have a tolerance issue with the adapter assembly, which could potentially not allow the snap ring to be seated correctly. If this is the case, there could be insufficient mount force that may cause the equipment to fall, resulting in serious injury.
Root cause
Component design/selection
Action
Stryker Communications sent an Urgent Medical Device Recall letter to all affected customers. A Stryker Representative will contact their facility to schedule a service visit. to replace the adapter assembly to resolve the tolerance issue affected units. Customers with questions were instructed to Stryker Technical Support at 800-243-5135 or comm.techservice@stryker.com. For questions regarding this recall call 972-834-8656.
Quantity
57 units
Distribution
Worldwide Distribution - US and Canada
Lot, serial or model codes
Serial Numbers: 7735170-X68377, 7735170-X68397, 7735170-X14846, 7735170-X15688, 7735170-X14492, 7735170-X14495, 7735170-X14497, 7735170-X14501, 7735170-X14502, 7735170-X14504, 7735170-X14506, 7735170-X14507, 7735170-X14508, 7735170-X14510, 7735170-X68383, 7735170-X68384, 7735170-X68385, 7735170-X68386, 7735170-X68393, 7735170-X68394, 7735170-X68395, 7735170-X68396, 7735170-X68392, 7735170-X68400, 7735170-X68401, 7735170-X15607, 7735170-X15608, 7735170-X15609, 7735170-X15610, 7735170-X15611, 7735170-X15612, 7735170-X15613, 7735170-X15614, 7735170-X15615
510(k) numbers
not on file
PMA numbers
not on file
Product code
KZF
Firm FEI
3002975342
Firm city
Flower Mound
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register