Skip to the content

FDA Medical Device Recalls (openFDA)

First Source Inc: First Source iQ Flex M. Portable cart and platform that supports an X-ray system and laptop computer.

Recall number
Z-2247-2024
Recalling firm
First Source Inc
Product
First Source iQ Flex M. Portable cart and platform that supports an X-ray system and laptop computer.
Status
Open, Classified
Initiated
2024-05-09
Posted
2024-07-01
Terminated
not on file
Reason
Potential for the bolt that secures the positioning arm to the gas spring to come loose and fall onto patient.
Root cause
Device Design
Action
Consignees were emailed an URGENT: MEDICAL DEVICE RECALL notification, dated 5/29/24. The notice instructs consignees to follow the recommended preventive maintenance schedule and visually check the device before use to identify if the affected bolt is coming loose. The recalling firm is currently testing a spring clip which will keep the bolt from falling out and will send this part to customers at no charge once approved. Consignees are asked to return the provided response form. Consignees with any questions can contact the firm by email at support@fsimed.com or by phone at 1-800-349-5980 (option 2) from 9:00 AM to 5:00 PM EST from Monday through Friday.
Quantity
137 units
Distribution
Worldwide - US Nationwide distribution in the states of AR, CA, FL, GA, IL, IN, LA, MA, MD, MI, MO, MS, NC, NY, OK, PA, TN, TX, & VA. The countries of Korea & Serbia.
Lot, serial or model codes
Model No. Pending; UDI-DI: G221500MC0740; Serial No. 4001, 4002, 4003, 4004, 4005, 4006, 4007, 4008, 4009, 4010, 4011, 4012, 4013, 4014, 4015, 4016, 4017, 4018, 4019, 4020, 4021, 4022, 4023, 4024, 4025, 4026, 4027, 4028, 4029, 4030, 4031, 4032, 4033, 4034, 4035, 4036, 4037, 4038, 4039, 4040, 4041, 4042, 4043, 4044, 4045, 4046, 4047, 4048, 4049, 4050, 4051, 4052, 4053, 4054, 4055, 4056, 4057, 4058, 4059, 4060.
510(k) numbers
["K221081"]
PMA numbers
not on file
Product code
IZL
Firm FEI
3004063527
Firm city
Rochester
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register