FDA Medical Device Recalls (openFDA)
Medtronic Sofamor Danek USA Inc: KYPHON(R) Directional Bone Filler Device, Size: 3, REF F04C, Rx Only, STERILE R.
- Recall number
- Z-2247-2017
- Recalling firm
- Medtronic Sofamor Danek USA Inc
- Product
- KYPHON(R) Directional Bone Filler Device, Size: 3, REF F04C, Rx Only, STERILE R.
- Status
- Terminated
- Initiated
- 2017-05-17
- Posted
- not on file
- Terminated
- 2019-06-18
- Reason
- The directional arrow at the proximal end of the instrument may not correctly align with the cutout opening on the distal end of the instrument. Using affected bone void filler may result in the injected cement being placed in a direction unintended by the surgeon.
- Root cause
- Nonconforming Material/Component
- Action
- Medtronic distributed a voluntary removal notice (Attachment I) via FedEx courier service to impacted U.S. consignees on 17 May 2017. Communications to impacted consignees outside the United States are currently ongoing per each geographies local procedures. The consignees were directed to do the following: 1.) Please locate and remove the impacted product from normal storage locations. Do not use this product. 2.) Complete the enclosed Customer Confirmation Form and return via fax to 763-367-8134, or via e-mail to rs.sdmriskmgt@medtronic.com. 3.) As applicable, contact your Medtronic Sales Representative to coordinate the return and replacement of affected product.
- Quantity
- 17657 units
- Distribution
- worldwide
- Lot, serial or model codes
- All Lots
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- OAR
- Firm FEI
- 1000477302
- Firm city
- Memphis
- Firm state
- TN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as