Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corporation: Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp - Extra Short. 1 Unit. Sterile. Baxter Healthcare Corporation, Deerfield, IL 60015 USA

Recall number
Z-2246-2017
Recalling firm
Baxter Healthcare Corporation
Product
Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp - Extra Short. 1 Unit. Sterile. Baxter Healthcare Corporation, Deerfield, IL 60015 USA
Status
Terminated
Initiated
2014-04-28
Posted
2017-06-02
Terminated
2017-07-21
Reason
Connection issues between the PD transfer set catheter connector and an adapter may lead to leaks and contamination of the fluid pathway which may result in peritonitis.
Root cause
Unknown/Undetermined by firm
Action
On April 28, 2014, firm sent an Urgent Product Recall to US customers via USPS 1st class mail. Baxter requested customers to locate and remove all affected product from their facility, and contact Baxter Healthcare Center for Service to arrange for return and credit. 100%, all non-respondents would be contacted by phone. All affected product would be destroyed. For general questions regarding this communication , contact The Center for One Baxter at 1-800-422-9837 , Monday through Friday, between the hours of 8:00AM and 5:00 PM Central Time.
Quantity
6,048
Distribution
Nationwide Distribution and Internationally to Canada and Malaysia
Lot, serial or model codes
Product Code 5C4483; Lot# H12030047
510(k) numbers
["K882498"]
PMA numbers
not on file
Product code
KDJ
Firm FEI
1417572
Firm city
Deerfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28