Skip to the content

FDA Medical Device Recalls (openFDA)

Ortho-Clinical Diagnostics, Inc.: VITROS Chemistry Products Calibrator Kit 29- In-Vitor used to calibrate the VITROS 5,1 FS/4600 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems for the quantitative measurement of total iron-binding capacity (TIBC) using VITROS Chemistry Products dTIBC Reagent Model/Catalog Number: 6802344

Recall number
Z-2245-2024
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
VITROS Chemistry Products Calibrator Kit 29- In-Vitor used to calibrate the VITROS 5,1 FS/4600 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems for the quantitative measurement of total iron-binding capacity (TIBC) using VITROS Chemistry Products dTIBC Reagent Model/Catalog Number: 6802344
Status
Open, Classified
Initiated
2024-05-14
Posted
2024-07-01
Terminated
not on file
Reason
Kit assigned an extended expiration date of 24-Jul-2025 during release. The correct expiration date for Lot 2953 is 24-Jul-2024. If used past its true expiration and the customer does not have their Assay Data Disk (ADD) updated, the system will allow them to use the expired calibrator which may bias the results. However, at this time the calibrator is within its use date and will function appropriately.
Root cause
Labeling design
Action
QuidelOrtho issued a hardcopy customer letters to Distributors and End-USers (Ref. CL2024-130) via FedEx overnight courier and/ or electronically using ORTHO PLUS e-Communications on 22 May 2024. Letter states reason for recall, health risk and action to take: Discontinue using, render unusable, and discard any remaining inventory of VITROS Chemistry Products Calibrator Kit 29, Lot 2953 by the correct expiration date, 24-Jul-2024. " If you have previously obtained a Certificate of Analysis, then download a new one from ORTHO PLUS on quidelortho.com. " To help assure the product is not supported beyond its expiry date, upload ADD DRV 6286 or higher before 24-Jul-2024. Once DRV 6286 or higher has been uploaded the product will no longer be supported and you must use an alternate lot. " Complete the enclosed Confirmation of Receipt form no later than May 29, 2024. Upon receipt of your completed Confirmation of Receipt form, QuidelOrtho will provide credit or replacement of your existing inventory of VITROS Chemistry Products Calibrator Kit 29, Lot 2953. " Save this notification with your User Documentation or post this notification by each VITROS 5,1 FS/4600/5600/XT 7600 System until the issue has been resolved. " Please forward this notification if the product was distributed outside of your facility. Contact Information. If you have further questions, please contact the Global Services Organization at 1-800-421-3311.
Quantity
1718
Distribution
Worldwide distribution - US Nationwide including Puerto Rico and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.
Lot, serial or model codes
(UDI/GTIN# 10758750002672; Lot Number 2953; Expiration Date: 24-Jul-2024
510(k) numbers
not on file
PMA numbers
not on file
Product code
JMO
Firm FEI
1000136573
Firm city
Rochester
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register