Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: Syngo.via, picture archiving and communications system software controlled. Intended to be used for viewing, manipulation, communication, and storage of medical images.

Recall number
Z-2245-2016
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Syngo.via, picture archiving and communications system software controlled. Intended to be used for viewing, manipulation, communication, and storage of medical images.
Status
Terminated
Initiated
2016-06-23
Posted
not on file
Terminated
2020-02-20
Reason
Incorrect values for the volume calculation. Software update VB30B via Update Instructions SY018/16/P to resolve software errors.
Root cause
Software design
Action
Siemens released and mailed to customers a notice with software update VB30B instructions via update instruction SY018/16/P for customers to follow until the update can be made. The letter also described the software issues involved; and asked customers to include the letter in the operator's manual until the update is performed. Customers are to ensure that all necessary personnel are aware of the situation.
Quantity
50 units
Distribution
Nationwide Distribution -- AR, CA, CO, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, ND, NE, NJ, NM, NV, NY, OH, OK, PA, SC, TX, and UT.
Lot, serial or model codes
310289,130542,130322,310149,130325,221063,310168,101762,221048,221096,220783,310195,130451,400031,130292,310114,220846,130269,130163,130402,130826,221051,102336,310282,220888,10007,221052,221049,130518,221002,221091,220915,130520,130472,100469,310249,130122,100013,130378,310188,102453,310079,310080,130240,130382,220985,310171,220879,220995, and 310268.
510(k) numbers
["K092519","K123375"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28