FDA Medical Device Recalls (openFDA)
Siemens Medical Solutions USA, Inc: Syngo.via, picture archiving and communications system software controlled. Intended to be used for viewing, manipulation, communication, and storage of medical images.
- Recall number
- Z-2245-2016
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Product
- Syngo.via, picture archiving and communications system software controlled. Intended to be used for viewing, manipulation, communication, and storage of medical images.
- Status
- Terminated
- Initiated
- 2016-06-23
- Posted
- not on file
- Terminated
- 2020-02-20
- Reason
- Incorrect values for the volume calculation. Software update VB30B via Update Instructions SY018/16/P to resolve software errors.
- Root cause
- Software design
- Action
- Siemens released and mailed to customers a notice with software update VB30B instructions via update instruction SY018/16/P for customers to follow until the update can be made. The letter also described the software issues involved; and asked customers to include the letter in the operator's manual until the update is performed. Customers are to ensure that all necessary personnel are aware of the situation.
- Quantity
- 50 units
- Distribution
- Nationwide Distribution -- AR, CA, CO, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, ND, NE, NJ, NM, NV, NY, OH, OK, PA, SC, TX, and UT.
- Lot, serial or model codes
- 310289,130542,130322,310149,130325,221063,310168,101762,221048,221096,220783,310195,130451,400031,130292,310114,220846,130269,130163,130402,130826,221051,102336,310282,220888,10007,221052,221049,130518,221002,221091,220915,130520,130472,100469,310249,130122,100013,130378,310188,102453,310079,310080,130240,130382,220985,310171,220879,220995, and 310268.
- 510(k) numbers
- ["K092519","K123375"]
- PMA numbers
- not on file
- Product code
- LLZ
- Firm FEI
- 2240869
- Firm city
- Malvern
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28